Laglera Luis M, Battaglia Gianluca, van den Berg Constant M G
Department of Earth and Ocean Sciences, University of Liverpool, Liverpool L69 3GP, UK.
Anal Chim Acta. 2007 Sep 5;599(1):58-66. doi: 10.1016/j.aca.2007.07.059. Epub 2007 Aug 1.
A new voltammetric method is presented for the measurement of humic substances (HS) in natural waters. The method is based on catalytic cathodic stripping voltammetry (CSV) and makes use of adsorptive properties of iron-HS complexes on the mercury drop electrode at natural pH. A fulvic acid standard (IHSS) was used to confirm the voltammetric response (peak potential and sensitivity) for the HS for natural water samples. Optimized conditions included the linear-sweep mode, deposition at -0.1 V, pH buffered at 8 and a scan rate of 50 mV s(-1). At a deposition time of 240 s in the presence of 10 nM iron and 30 mM bromate, the detection limit was 5 microg L(-1) HS in seawater, which could be lowered further by an increase in the bromate concentration, or in the adsorption time. The method was used to determine HS in the Irish Sea which were found to occur at levels between 60 and 600 microg L(-1). The new method is sufficiently sensitive to detect the low HS content in oceanic samples and has implications to the study of iron speciation.
提出了一种用于测量天然水中腐殖质(HS)的新伏安法。该方法基于催化阴极溶出伏安法(CSV),利用铁 - HS络合物在自然pH下于汞滴电极上的吸附特性。使用富里酸标准品(国际腐殖质学会标准品)来确认天然水样中HS的伏安响应(峰电位和灵敏度)。优化条件包括线性扫描模式、在 - 0.1 V下沉积、pH缓冲至8以及扫描速率为50 mV s⁻¹。在10 nM铁和30 mM溴酸盐存在下,沉积时间为240 s时,海水中HS的检测限为5 μg L⁻¹,通过增加溴酸盐浓度或吸附时间可进一步降低检测限。该方法用于测定爱尔兰海中的HS,发现其含量在60至600 μg L⁻¹之间。新方法灵敏度足以检测海洋样品中低含量的HS,对铁形态研究具有重要意义。