Limitations of rate response of an activity-sensing rate-responsive pacemaker to different forms of activity.

作者信息

Lau C P, Mehta D, Toff W D, Stott R J, Ward D E, Camm A J

机构信息

Department of Cardiovascular Sciences, St. George's Hospital, London, England.

出版信息

Pacing Clin Electrophysiol. 1988 Feb;11(2):141-50. doi: 10.1111/j.1540-8159.1988.tb04535.x.

Abstract

The responses of an activity-sensing rate-responsive system (Activitrax) to various forms of physiological activity were assessed in 15 individuals who had this pacemaker. Nine were patients with complete heart block and atrial arrhythmias; their mean age was 60 years (range, 41-85 years). Six were age-matched healthy volunteers who were exercised with an external Activitrax system attached firmly to the chest wall. The pacemaker was programmed to achieve a pacing rate of about 100 bpm at the end of the first stage of the Bruce protocol (pacemaker settings: rate = 70-150 bpm; threshold = low to medium; response = 6-9). In the activity-sensing ventricular pacing mode, all patients achieved a significant increase in treadmill time compared to constant-rate ventricular pacing (mean +/- SD, 8.0 +/- 3.3 vs 5.4 +/- 2.3 minutes; p less than 0.01), with a mean maximum pacing rate of 123 +/- 18 bpm. Jogging in place produced a prompt increase in pacing rate, with the maximum achieved at the end of the exercise. However, physiological activities such as hand-grip, the Valsalva maneuver and standing resulted in only minimal rate response. Pacing rate after ascending 4 flights of stairs was the same as that achieved after descending the same stairs (100 +/- 8 vs 105 +/- 4 bpm; p = 0.1). All 15 subjects were exercised from resting heart rate for 3 minutes on a treadmill at 1.2 mph and 2.5 mph with four gradients at each speed. Although the pacing rate increased with a faster treadmill speed (p less than 0.005), it did not respond appropriately to a change in gradient compared to the sinus rate. We conclude that although activity-sensing rate-responsive pacing gives a prompt increase in pacing rate and improves maximum exercise tolerance, further refinement is necessary because: (1) physiological activities not associated with significant movement are not detected by this pacing system; (2) detection of vibrations as an indicator of activities does not correlate well with the level of exertion.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验