Suppr超能文献

运动时犬心内膜和心外膜心电图电压降低:对起搏器感知的影响。

Decrease in canine endocardial and epicardial electrogram voltages with exercise: implications for pacemaker sensing.

作者信息

Bricker J T, Ward K A, Zinner A, Gillette P C

机构信息

Lillie Frank Abercrombie Section of Cardiology, Baylor College of Medicine, Houston, Texas.

出版信息

Pacing Clin Electrophysiol. 1988 Apr;11(4):460-4. doi: 10.1111/j.1540-8159.1988.tb06007.x.

Abstract

Individuals with permanently implanted pacemakers who have normal sensing and pacing at rest may demonstrate abnormalities, particularly of atrial sensing, with exercise. Exercise is known to cause changes in the surface electrocardiographic voltages. The purpose of this study was to evaluate changes in endocardial and epicardial voltages during exercise in canines with permanently implanted pacing leads similar to those permanently implanted in humans with pacemakers. Six adult mongrel dogs were equipped with transvenous atrial and ventricular endocardial bipolar screw-in leads from a jugular venosection, and with atrial and ventricular bipolar stab-on leads from a left thoracotomy. Exercise was performed at one week following implantation and at weekly intervals for a total of 4 weeks. The animals were studied at rest and at a single workload of 3 miles per hour at a 30% elevation. Unfiltered electrograms were recorded with a VR-12 Electronics-for-Medicine photographic recorder. The mean decreases of electrograms with exercise from resting voltage were: 15% for atrial endocardial unipolar leads; 11% for atrial endocardial bipolar leads; 4% for atrial epicardial unipolar leads; 15% for atrial epicardial bipolar leads; 8% for ventricular endocardial unipolar leads; 18% for ventricular endocardial bipolar leads; 0.1% for ventricular epicardial unipolar leads and 5% for ventricular epicardial bipolar leads.

摘要

对于植入永久性起搏器且静息时感知和起搏功能正常的个体,运动时可能会出现异常,尤其是心房感知异常。已知运动可导致体表心电图电压发生变化。本研究的目的是评估在犬类中,与人类永久性植入起搏器类似的永久性植入起搏导线在运动过程中心内膜和心外膜电压的变化。六只成年杂种犬通过颈静脉切开术植入经静脉心房和心室心内膜双极螺旋电极,通过左胸切开术植入心房和心室双极穿刺电极。在植入后一周进行运动,并每周进行一次,共持续4周。在静息状态和以每小时3英里、坡度30%的单一工作量运动时对动物进行研究。使用VR-12医用电子摄影记录仪记录未滤波的心电图。运动时心电图相对于静息电压的平均下降幅度为:心房心内膜单极电极15%;心房心内膜双极电极11%;心房心外膜单极电极4%;心房心外膜双极电极15%;心室心内膜单极电极8%;心室心内膜双极电极18%;心室心外膜单极电极0.1%;心室心外膜双极电极5%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验