Anspach B, Gierlich H U, Unger K K
Institut für Anorganische Chemie und Analytische Chemie, Johannes Gutenberg Universität, Mainz, F.R.G.
J Chromatogr. 1988 Jun 29;443:45-54. doi: 10.1016/s0021-9673(00)94781-2.
A reduction of the mean particle diameter of silica-based packings in the size-exclusion chromatography (SEC) of proteins to about 5 micron generates the expected increase in column plate number over the traditional 10 micron SEC columns, as demonstrated for the Zorbax Bio Series GF 250 and GF 450 and TSK-Gel 3000 SWXL columns. The slightly lower column efficiency of the TSK-Gel 3000 SWXL compared with the GF 250 column is compensated by the fact that the phase ratio of the 3000 SWXL column is higher by a factor of two. Hence both columns show nearly the same peak capacity of about 20-30. When the ionic strength of the eluent was changed by varying the salt concentration, the elution volume of proteins was found to be affected differently on the two columns. These alterations are attributed to the differences in the surface compositions of the two types of packings.
在蛋白质的尺寸排阻色谱法(SEC)中,将基于硅胶的填料的平均粒径减小至约5微米,与传统的10微米SEC柱相比,柱板数会如预期那样增加,这在Zorbax Bio系列GF 250和GF 450以及TSK-Gel 3000 SWXL柱上得到了证明。与GF 250柱相比,TSK-Gel 3000 SWXL柱的柱效略低,但3000 SWXL柱的相比率高出两倍,弥补了这一不足。因此,两根柱子的峰容量几乎相同,约为20 - 30。当通过改变盐浓度来改变洗脱液的离子强度时,发现蛋白质在两根柱子上的洗脱体积受到的影响不同。这些变化归因于两种填料表面组成的差异。