Marco D D, Gallagher D
Siemens-Pacesetter, Inc., Sylmar, CA.
Pacing Clin Electrophysiol. 1988 Nov;11(11 Pt 2):1673-8. doi: 10.1111/j.1540-8159.1988.tb06293.x.
Retrograde, or ventriculoatrial conduction (VAC) has been measured in 67% of pacemaker patients with SSS and 25% of patients with 3 degrees block. DDD pacers use the postventricular atrial refractory period (PVARP) to prevent sensing VAC. If the PVARP is too short, a pacer-mediated tachycardia (PMT) could occur. If too long, rapid atrial rates may not be sensed. To accurately set the PVARP, VAC must be measured. Five noninvasive methods are described: (1) Surface ECG during VVI pacing; (2) Atrial intracardiac electrograms during VVI pacing; (3) Initiating a PMT with long pacemaker AV interval; (4) Initiating a PMT at the maximum tracking rate; (5) DDI mode with electronic calipers. The description of these methods is intended to enable a physician to detect and measure VAC in patients with any manufacturer's DDD or VDD pacemaker.
在67%的病态窦房结综合征起搏器患者和25%的三度房室传导阻滞患者中检测到了逆行性或心室心房传导(VAC)。DDD起搏器利用心室后心房不应期(PVARP)来防止感知VAC。如果PVARP过短,可能会发生起搏器介导的心动过速(PMT)。如果过长,则可能无法感知快速的心房率。为了准确设置PVARP,必须测量VAC。文中描述了五种非侵入性方法:(1)VVI起搏时的体表心电图;(2)VVI起搏时的心内房内电图;(3)通过长起搏器房室间期引发PMT;(4)以最大跟踪速率引发PMT;(5)使用电子卡尺的DDI模式。对这些方法的描述旨在使医生能够在使用任何制造商的DDD或VDD起搏器的患者中检测和测量VAC。