Clinical experience with steroid-eluting unipolar electrodes.

作者信息

Pirzada F A, Moschitto L J, Diorio D

机构信息

Department of Cardiology, Malden Hospital, Boston University School of Medicine, Massachusetts 02148.

出版信息

Pacing Clin Electrophysiol. 1988 Nov;11(11 Pt 2):1739-44. doi: 10.1111/j.1540-8159.1988.tb06304.x.

Abstract

In continuing search of low chronic threshold leads, a new concept of electrode design which is capable of delivering corticosteroids at the myocardial tissue interface has been made available by Medtronic. Twenty-three patients, 17 females and 6 males, were either implanted with 4003 (n = 21) or 5023 (n = 2) steroid-eluting electrodes in the ventricular chamber. Pacing modes utilized were VVIM (n = 13) or DDD (n = 10). Pulse generators used were Medtronic (7005, 8317, 8329) Pacesetter (285) and Intermedics (283). Thresholds at the time of implantation at 0.50 msec pulse width were 0.40 +/- 0.02 volts at 0.66 +/- 0.05 milliamps. Resistance and R wave measured were 565.43 +/- 22.07 ohms and 9.24 +/- 1.06 mv, respectively. Chronic thresholds were checked on routine follow-up visits by either decreasing pulse width and/or pulse amplitude. Data is being reported between 1 and 88 (23.22 +/- 4.35) weeks. Pulse width threshold at 2.5 volts were 0.10 msec (n = 11) and 0.05 msec or lower (n = 12). At 5.0 volts no loss of capture was seen at 0.05 msec (n = 22) except in one patient at 0.10 msec. Pulse width thresholds in the first 24 weeks were lower than 0.20 msec at 2.5 volts (n = 15) and less than 0.70 msec. at 0.8 volts (n = 6). No loss of sensing was seen by electrocardiographic analysis at the time of threshold checks with the pulse generator at standard setting of the R wave. Thus, in this initial report, the steroid-eluting electrodes have demonstrated very low thresholds both in the early and chronic follow-up phase.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

为了持续寻找低慢性阈值导线,美敦力公司推出了一种新的电极设计概念,该电极能够在心肌组织界面释放皮质类固醇。23例患者,17例女性和6例男性,在心室植入了4003型(n = 21)或5023型(n = 2)类固醇洗脱电极。使用的起搏模式为VVIM(n = 13)或DDD(n = 10)。使用的脉冲发生器有美敦力(7005、8317、8329)、百胜(285)和英特迈迪克(283)。植入时脉宽为0.50毫秒时的阈值为0.40±0.02伏,电流为0.66±0.05毫安。测得的电阻和R波分别为565.43±22.07欧姆和9.24±1.06毫伏。在常规随访中,通过减小脉宽和/或脉冲幅度来检查慢性阈值。报告的数据为1至88(23.22±4.35)周。2.5伏时的脉宽阈值为0.10毫秒(n = 11)和0.05毫秒或更低(n = 12)。在5.0伏时,除1例患者脉宽为0.10毫秒外,其余22例患者在0.05毫秒时均未出现夺获丧失。在最初24周内,2.5伏时脉宽阈值低于0.20毫秒(n = 15),0.8伏时低于0.70毫秒(n = 6)。在阈值检查时,将脉冲发生器设置为R波的标准设置,通过心电图分析未发现感知丧失。因此,在这份初步报告中,类固醇洗脱电极在早期和慢性随访阶段均显示出非常低的阈值。(摘要截取自250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验