Suppr超能文献

较小电极表面积(4平方毫米)对类固醇洗脱导线的益处。

Benefits of smaller electrode surface area (4 mm2) on steroid-eluting leads.

作者信息

Schuchert A, Kuck K H

机构信息

Department of Cardiology, University-Hospital Eppendorf, Hamburg, Federal Republic of Germany.

出版信息

Pacing Clin Electrophysiol. 1991 Dec;14(12):2098-104. doi: 10.1111/j.1540-8159.1991.tb06480.x.

Abstract

The purpose was to test whether a reduction of pacemaker electrode surface area below 8 mm2 improves leads that elute steroid from the electrode tip to the surrounding myocardium. A standard-sized 8 mm2 lead with 1 mg dexamethasone was implanted in 12 patients and a lead with 4 mm2 electrode surface area and 0.5 mg dexamethasone in ten patients. Pacing threshold, impedance, and sensing threshold were measured at implantation and after 1, 4, and 12 weeks. Pacing thresholds were similar for both groups and were always less than or equal to 0.8 V at 0.5 msec pulse duration in all patients. Impedance was significantly higher (P less than 0.05) for the 4 mm2 lead (implantation: 726 +/- 119 ohms; 1 week: 596 +/- 71 ohms; 4 weeks: 624 +/- 68 ohms; 12 weeks: 643 +/- 56 ohms) than for the 8 mm2 lead (implantation: 422 +/- 43 ohms; 1 week: 402 +/- 48 ohms; 4 weeks: 439 +/- 57 ohms; 12 weeks: 449 +/- 61 ohms). R wave amplitudes did not differ between both groups; no sensing failure occurred at 5 mV sensitivity. Compared to the 8 mm2 lead the reduction of surface area to 4 mm2 did not influence pacing threshold, but resulted in a higher pacing impedance. The amount of pacing energy was lower in the smaller-sized electrode. For clinical impact, low pacing threshold and high impedance leads are the condition to implant pulse generators with smaller battery capacity.

摘要

目的是测试将起搏器电极表面积减小至8平方毫米以下是否能改善从电极尖端向周围心肌洗脱类固醇的导线。将一根带有1毫克地塞米松的标准尺寸8平方毫米导线植入12例患者体内,将一根电极表面积为4平方毫米且带有0.5毫克地塞米松的导线植入10例患者体内。在植入时以及1周、4周和12周后测量起搏阈值、阻抗和感知阈值。两组的起搏阈值相似,在所有患者中,脉宽为0.5毫秒时起搏阈值始终小于或等于0.8伏。4平方毫米导线的阻抗(植入时:726±119欧姆;1周时:596±71欧姆;4周时:624±68欧姆;12周时:643±56欧姆)显著高于8平方毫米导线(植入时:422±43欧姆;1周时:402±48欧姆;4周时:439±57欧姆;12周时:449±61欧姆)(P<0.05)。两组之间的R波振幅无差异;在5毫伏灵敏度下未发生感知失败。与8平方毫米导线相比,表面积减小至4平方毫米对起搏阈值无影响,但导致起搏阻抗更高。较小尺寸电极的起搏能量更低。对于临床影响而言,低起搏阈值和高阻抗导线是植入电池容量较小的脉冲发生器的条件。

相似文献

3
Initial experience with 1.5-mm2 high impedance, steroid-eluting pacing electrodes.
Pacing Clin Electrophysiol. 1996 Feb;19(2):188-96. doi: 10.1111/j.1540-8159.1996.tb03310.x.
6
Evolution of an active fixation atrial pacing lead.主动固定心房起搏导线的演变
Pacing Clin Electrophysiol. 2000 Nov;23(11 Pt 2):1795-7. doi: 10.1111/j.1540-8159.2000.tb07021.x.
8
A new efficient NanoTip lead.一种新型高效纳米尖端导线。
Pacing Clin Electrophysiol. 1990 Dec;13(12 Pt 2):1901-5. doi: 10.1111/j.1540-8159.1990.tb06913.x.
10
Paradigm shift in lead design.先导物设计中的范式转变。
Pacing Clin Electrophysiol. 1999 Sep;22(9):1321-32. doi: 10.1111/j.1540-8159.1999.tb00625.x.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验