Mercando A D, Furman S, Johnston D, Frame R, Brodman R, Kim S G, Fisher J D
Department of Medicine, Montefiore Medical Center, Bronx, New York 10467.
Pacing Clin Electrophysiol. 1988 Nov;11(11 Pt 2):2059-63. doi: 10.1111/j.1540-8159.1988.tb06350.x.
Between May 1982 and May 1988, 37 patients (28 males and 9 females, mean age 57.6, range 16-76 years) of approximately 600 evaluated for sustained ventricular tachycardia and/or fibrillation (VT/VF) were treated with an automatic implantable cardioverter defibrillator (AICD). Twenty-eight of the patients had coronary artery disease, 7 had nonischemic cardiomyopathy, 1 had amyloid heart disease, and 1 had rheumatic heart disease. The mean ejection fraction was 32.2 +/- 12.9% (range, 9-64%). Eleven patients have died at a mean of 16.7 months after implantation. The cumulative survival rate was 81% at 1 year, 77% at 2 years, 68% at 3 years, and 53% at 4, 5, and 6 years. Considering only sudden deaths, the survival was 97% at 1 and 2 years, 90% at 3 years, and 80% at 4, 5, and 6 years. Twenty-one of the 37 patients received spontaneous shocks. If the first shock marks the time to death in the absence of an AICD, the cumulative survival rate would have been 56% at 1 year, 42% at 2 years, 29% at 3 years, and 14% at 4, 5, and 6 years. The maximum amount of time to a first appropriate shock was 39.7 months. Thirty-nine devices have been explanted: 28 for battery depletion; 5 for infections; 3 for improper sensing; 2 for electronic failure; and 1 at the time of cardiac transplantation. The average time to failure of the 28 units removed for battery depletion was 19.8 +/- 6.9 months.(ABSTRACT TRUNCATED AT 250 WORDS)
1982年5月至1988年5月期间,在约600例因持续性室性心动过速和/或心室颤动(VT/VF)接受评估的患者中,37例(28例男性和9例女性,平均年龄57.6岁,范围16 - 76岁)接受了植入式自动心脏复律除颤器(AICD)治疗。其中28例患者患有冠状动脉疾病,7例患有非缺血性心肌病,1例患有淀粉样心脏病,1例患有风湿性心脏病。平均射血分数为32.2±12.9%(范围9 - 64%)。11例患者在植入后平均16.7个月死亡。1年时累积生存率为81%,2年时为77%,3年时为68%,4、5和6年时为53%。仅考虑猝死情况,1年和2年时生存率为97%,3年时为90%,4、5和6年时为80%。37例患者中有21例接受了自发除颤。如果在没有AICD的情况下,首次除颤标志着死亡时间,那么1年时累积生存率将为56%,2年时为42%,3年时为29%,4、5和6年时为14%。首次适当除颤的最长时间为39.7个月。已取出39台设备:28台因电池耗尽;5台因感染;3台因感知不当;2台因电子故障;1台在心脏移植时取出。因电池耗尽取出的28台设备的平均故障时间为19.8±6.9个月。(摘要截选至250字)