Eastep S J, Benson P J, Preese L M, Apple F S
Clinical Laboratories #812, Hennepin County Medical Center, Minneapolis, MN 55415.
Clin Chem. 1989 Feb;35(2):333-4.
Sera from patients with above-normal gamma-globulin concentrations give sodium values as much as 10% higher when measured in the Kodak Ektachem 400 analyzer than results obtained with the Beckman Astra-8 analyzer. Results obtained with the Astra were in agreement with flame photometry. A similar relative bias was observed for potassium measurement in the Ektachem. This discrepancy was corrected when we used a new generation of reference fluid provided by Kodak, which is currently available for general use.