偏瘫患者平衡评估中压力中心测量的日内可靠性。

Between-day reliability of centre of pressure measures for balance assessment in hemiplegic stroke patients.

机构信息

Explorations Fonctionnelles Physiologiques, CHU Toulouse Rangueil, Avenue du Pr Jean Poulhès, 31059 Toulouse, France.

出版信息

J Neuroeng Rehabil. 2014 Mar 21;11:39. doi: 10.1186/1743-0003-11-39.

Abstract

BACKGROUND

Stroke patients have impaired postural balance that increases the risk of falls and impairs their mobility. Assessment of postural balance is commonly carried out by recording centre of pressure (CoP) displacements, but the lack of data concerning reliability of these measures compromises their interpretation. The purpose of this study was to investigate the between-day reliability of six CoP-based variables, in order to provide i) reliability data for monitoring postural sway and weight-bearing asymmetry of stroke patients in clinical practice and ii) consistent assessment method of measurement error for applications in physical medicine and rehabilitation.

METHODS

Postural balance of 20 stroke patients was assessed in quiet standing on a force platform, in two sessions, 7 days apart. Six CoP-based variables were collected in eyes open and eyes closed conditions: postural sway was assessed with mean and standart deviation of CoP-velocity, CoP-velocity along the mediolateral and anteroposterior axes, and confidence ellipse area (CE(AREA)); weight-bearing asymmetry was assessed with mean CoP position along the mediolateral axis (CoP(ML)). The intraclass correlation coefficient (ICC) was used to determine the level of agreement between test-retest. Small real difference (SRD), corresponding to the smallest change that indicates a real improvement for a single individual, was used to determine the extent of measurement error.

RESULTS

ICCs were satisfactory (>0.9) for all CoP-based variables, except for CE(AREA) in eyes open condition and CoP(ML) (<0.8). The SRDs (eyes open/closed conditions) were: 6.1/9.5 mm.s(-1) for mean velocity; 12.3/12.2 mm.s(-1) for standard deviation of CoP-velocity; 3.6/5.5 mm.s(-1) and 4.9/7.3 mm.s(-1) for CoP-velocity in mediolateral and anteroposterior axes, respectively; 17.4/21.4 mm for CoP(ML). Because CE(AREA) showed heteroscedasticity of measurement error distribution, SRD (eyes open/closed conditions) was expressed as a percentage (121/75%) and a ratio (3.68/2.16) obtained after log-antilog procedure.

CONCLUSIONS

In clinical practice, the CoP-based velocity variables should be prefer to CE(AREA) to assess and monitor postural sway over time in hemiplegic stroke patients. The poor reliability of CoP(ML) compromises its use to assess weight-bearing asymmetry. The procedure we used could be applied in reliability studies concerning other CoP-based variables or other biological variables in the field of physical medicine and rehabilitation.

摘要

背景

中风患者的姿势平衡受损,增加了跌倒的风险,并损害了他们的活动能力。姿势平衡的评估通常通过记录中心压力(CoP)位移来进行,但这些测量的可靠性数据缺乏,影响了其解释。本研究的目的是探讨六种基于 CoP 的变量的日内可靠性,以便为监测中风患者的姿势摆动和承重不对称性提供:i)在临床实践中,ii)为物理医学和康复领域的测量误差提供一致的评估方法。

方法

在为期 7 天的两次会议中,20 名中风患者在力平台上睁眼和闭眼安静站立时进行姿势平衡评估。在睁眼和闭眼条件下收集了 6 种基于 CoP 的变量:姿势摆动用 CoP-速度的平均值和标准差、CoP-速度在前后和左右方向上的平均值和标准差,以及置信椭圆面积(CE(AREA));承重不对称性用 CoP 在左右轴上的平均位置(CoP(ML))评估。使用组内相关系数(ICC)来确定测试-再测试之间的一致性水平。小实际差异(SRD),对应于单个个体实际改善的最小变化,用于确定测量误差的程度。

结果

除了睁眼条件下的 CE(AREA)和 CoP(ML)(<0.8)外,所有基于 CoP 的变量的 ICC 均令人满意(>0.9)。SRD(睁眼/闭眼条件)分别为:6.1/9.5mm.s(-1)的平均速度;12.3/12.2mm.s(-1)的 CoP-速度标准差;3.6/5.5mm.s(-1)和 4.9/7.3mm.s(-1)的 CoP-速度在左右和前后方向上,分别为 17.4/21.4mm 的 CoP(ML)。由于 CE(AREA)显示出测量误差分布的异方差性,SRD(睁眼/闭眼条件)表示为百分比(121/75%)和对数-反对数处理后获得的比值(3.68/2.16)。

结论

在临床实践中,与 CE(AREA)相比,偏瘫中风患者应优先选择基于 CoP 的速度变量来评估和监测随时间推移的姿势摆动。CoP(ML)的可靠性较差,影响了其对承重不对称性的评估。我们使用的程序可应用于物理医学和康复领域的其他基于 CoP 的变量或其他生物变量的可靠性研究。