Badiou Stéphanie, Guillot Julie, Kuster Nils, Bargnoux Anne-Sophie, Aguilar-Martinez Patricia, Boissier Elodie, Cristol Jean-Paul, Dupuy Anne-Marie
Department of Biochemistry, Montpellier University Hospital, Montpellier, France; UMR 204 NUTRIPASS, Université Montpellier I/II, Montpellier, France.
J Clin Lab Anal. 2014 Nov;28(6):428-34. doi: 10.1002/jcla.21705. Epub 2014 Mar 20.
This study aimed to evaluate a new ion-exchange HPLC system to measure hemoglobin A1c (HbA1c) in comparison to two other widely used HPLC systems.
Analytical performance and hemoglobin variants detection was assessed on the new Arkray/ELITech ADAMS HA-8180V, in parallel to Bio-Rad Variant II Turbo2.0 and Tosoh Bioscience HLC-723G8. A method comparison and concordance for the classification of patients according to the American Diabetes Association (ADA) categories was performed using kappa test.
ADAMS HA-8180V demonstrated excellent within-run imprecision (0%) and between-run imprecision: ≤1.21% using blood sample and ≤0.94% using quality controls. Method comparison of ADAMS HA-8180V with the two other systems yielded very high coefficient correlation (r > 0.995). A very good agreement (kappa value ≥0.81) was observed between methods for the classification of patients according to the ADA categories. In addition, all systems were able to detect the presence of most classical variants such as HbC, HbS, HbD, and HbE.
The Arkray/ELITech ADAMS HA-8180V demonstrated high analytical performance similar to previous systems such as Biorad(TM) Variant II Turbo 2.0 and Tosoh Bioscience HLC-723G8. The three systems allow a high-quality HbA1c measurement and appeared interchangeable for diabetes diagnosis or for the therapeutic monitoring of patients without hemoglobin variants.
本研究旨在评估一种新型离子交换高效液相色谱系统,以测量糖化血红蛋白(HbA1c),并与另外两种广泛使用的高效液相色谱系统进行比较。
在新的Arkray/ELITech ADAMS HA - 8180V上评估分析性能和血红蛋白变异体检测,同时与Bio - Rad Variant II Turbo2.0和Tosoh Bioscience HLC - 723G8进行比较。使用kappa检验对根据美国糖尿病协会(ADA)类别对患者进行分类的方法进行比较和一致性分析。
ADAMS HA - 8180V显示出出色的批内不精密度(0%)和批间不精密度:使用血样时≤1.21%,使用质量控制品时≤0.94%。ADAMS HA - 8180V与其他两个系统的方法比较产生了非常高的系数相关性(r > 0.995)。在根据ADA类别对患者进行分类的方法之间观察到非常好的一致性(kappa值≥0.81)。此外,所有系统都能够检测到大多数经典变异体的存在,如HbC、HbS、HbD和HbE。
Arkray/ELITech ADAMS HA - 8180V显示出与先前系统如Biorad™ Variant II Turbo 2.0和Tosoh Bioscience HLC - 723G8相似的高分析性能。这三种系统允许进行高质量的HbA1c测量,并且对于无血红蛋白变异体的糖尿病诊断或患者治疗监测似乎是可互换的。