Suppr超能文献

一种用于绑扎主肺动脉的假体,能够通过球囊血管成形术进行系列扩张。

A prosthesis for banding the main pulmonary artery, capable of serial dilatation by balloon angioplasty.

作者信息

Vince D J, Culham J A

机构信息

University of British Columbia, British Columbia Children's Hospital, Vancouver, Canada.

出版信息

J Thorac Cardiovasc Surg. 1989 Mar;97(3):421-7.

PMID:2465458
Abstract

A prosthesis constructed with a fatigued steel helix encased in a silicone rubber shield was used to band the main pulmonary artery in 10 dogs. After a mean duration of 138 days the banded site was dilated with a 20 mm diameter angioplasty catheter. This dilatation produced a mean increase of 44.3% in the cross-sectional area. A further mean increase of 2.2% in the cross-sectional area was measured 137 days after the dilatation. In five uncomplicated experiments a second dilatation was performed with a 23 mm diameter angioplasty catheter after a mean interval of 140 days. The second dilatation produced a further 21% increase in the cross-sectional area. In the five experiments in which two dilatations were performed, there was a total increase in the mean cross-sectional area of 94% produced 273 days after banding. This prosthesis maintains banding of the main pulmonary artery and can be serially dilated by balloon angioplasty.

摘要

用包裹在硅橡胶护套中的疲劳钢螺旋制成的假体对10只狗的主肺动脉进行束带。平均138天后,用直径20毫米的血管成形术导管扩张束带部位。这种扩张使横截面积平均增加了44.3%。扩张后137天,横截面积又平均增加了2.2%。在5个无并发症的实验中,平均间隔140天后,用直径23毫米的血管成形术导管进行第二次扩张。第二次扩张使横截面积又增加了21%。在进行了两次扩张的5个实验中,束带后273天平均横截面积总共增加了94%。这种假体可维持主肺动脉的束带状态,并可通过球囊血管成形术进行连续扩张。