Vince D J, Culham J A, Taylor G P
Radiology. 1987 Jul;164(1):141-4. doi: 10.1148/radiology.164.1.2954182.
Two arterial banding prostheses were developed that could be dilated by an intraluminal balloon dilator. One prosthesis consisted of radiopaque umbilical tape with a pleat secured by four 8-0 silk sutures thinly coated with silicone type A medical adhesive. This design was used to band the main pulmonary artery in five dogs, the subclavian artery in five, and the aorta in one. The second prosthesis, a stainless steel fatigued helix encased in a siliconized shield, was used to band the main pulmonary artery in two dogs and the descending aorta in two. After a mean duration of 89 days the bands were dilated with an intraluminal balloon dilator at 6 atm (608 kPa) of pressure for 30 seconds. In all 15 experiments the bands dilated and the gradient at the band was reduced. The stainless steel helix was more successful and has technical advantages for clinical application.