新型圣犹达Trifecta主动脉生物假体运动期间的血流动力学性能:一项法国多中心研究的结果。

Hemodynamic performance during exercise of the new St. Jude Trifecta aortic bioprosthesis: results from a French multicenter study.

作者信息

Levy Franck, Donal Erwan, Bière Loïc, Szymanski Catherine, Remadi Jean Paul, Flécher Erwan, Fouquet Olivier, Leguerrier Alain, Tribouilloy Christophe

机构信息

Department of Cardiology and INSERM and ERI-12, University Hospital, Amiens, France.

CIC-IT U 804, CHU Rennes, Université Rennes 1, Rennes, France; Service de Cardiologie, CHU Rennes, Rennes, France; INSERM, U1099, Rennes, France; Université de Rennes 1, LTSI, Rennes, France.

出版信息

J Am Soc Echocardiogr. 2014 Jun;27(6):590-7. doi: 10.1016/j.echo.2014.01.022. Epub 2014 Mar 18.

Abstract

BACKGROUND

Initial experience with the new St. Jude Trifecta pericardial aortic stented bioprosthesis shows an excellent resting hemodynamic profile. Little is known about changes in the hemodynamic profile of the Trifecta valve during exercise.

METHODS

Between February 2011 and November 2012, 85 patients (49 men; mean age, 76 ± 7 years) with severe symptomatic aortic stenosis who underwent aortic valve replacement with the Trifecta bioprosthesis at three centers in France (Amiens, Rennes, and Angers) underwent quantitative Doppler echocardiographic at rest, during low-level exercise (25 W), and during peak exercise (68 ± 21 W), 6 months after aortic valve replacement.

RESULTS

Mean peak transvalvular aortic velocity, mean transvalvular gradient, and mean left ventricular ejection fraction for all valve sizes were 211 ± 35 cm/sec, 10 ± 3 mm Hg, and 62 ± 10% at rest; 237 ± 48 cm/sec, 13 ± 4 mm Hg, and 64 ± 10% during low-level exercise; and 248 ± 70 cm/sec, 15 ± 5 mm Hg, and 67 ± 10% during peak exercise, respectively. Mean effective orifice area was 1.84 ± 0.42 cm(2) at rest, 1.86 ± 0.84 cm(2) (P = .92) during low-level exercise, and 1.95 ± 0.62 cm(2) (P = .49) during peak exercise. The prevalence of prosthesis-patient mismatch was low in the overall series (23%) and increased to 30% for the smallest valve sizes (19 and 21 mm).

CONCLUSIONS

The new Trifecta bioprosthesis provides an excellent hemodynamic profile both at rest and during exercise. This type of valve could be an appropriate choice in patients with small aortic annular diameters, to avoid prosthesis-patient mismatch.

摘要

背景

新型圣犹达Trifecta心包主动脉带支架生物瓣膜的初步使用经验显示其静息血流动力学特征极佳。关于Trifecta瓣膜在运动期间血流动力学特征的变化知之甚少。

方法

在2011年2月至2012年11月期间,法国三个中心(亚眠、雷恩和昂热)的85例(49例男性;平均年龄76±7岁)有症状的重度主动脉瓣狭窄患者,在主动脉瓣置换术后6个月,接受了Trifecta生物瓣膜置换,并分别于静息状态、低水平运动(25瓦)和峰值运动(68±21瓦)时进行了定量多普勒超声心动图检查。

结果

所有瓣膜尺寸的平均跨瓣主动脉峰值流速、平均跨瓣压差和平均左心室射血分数在静息时分别为211±35厘米/秒、10±3毫米汞柱和62±10%;在低水平运动时分别为237±48厘米/秒、13±4毫米汞柱和64±10%;在峰值运动时分别为248±70厘米/秒、15±5毫米汞柱和67±10%。平均有效瓣口面积在静息时为1.84±0.42平方厘米,在低水平运动时为1.86±0.84平方厘米(P = 0.92),在峰值运动时为1.95±0.62平方厘米(P = 0.49)。在整个系列中,人工瓣膜-患者不匹配的发生率较低(23%),对于最小的瓣膜尺寸(19和21毫米),该发生率增加到30%。

结论

新型Trifecta生物瓣膜在静息和运动时均提供了极佳的血流动力学特征。对于主动脉瓣环直径较小的患者,这种类型的瓣膜可能是一个合适的选择,以避免人工瓣膜-患者不匹配。