Tasca Giordano, Redaelli Paola, Riva Beatrice, De Carlini Caterina Chiara, Lobiati Elisabetta, Gamba Amando
Department of Cardiovascular, Operative Unit of Cardiac Surgery, Ospedale "A. Manzoni" di Lecco, Lecco, Italy; Department of Electronic, Information and Bioengeering, Politecnico di Milano, Milan, Italy.
J Card Surg. 2015 May;30(5):400-4. doi: 10.1111/jocs.12536. Epub 2015 Mar 10.
Patients with a small aortic annulus, that is ≤ 23 mm, constitute a challenge for the surgeon, because they are at high risk of patient-prosthesis mismatch. Stentless valves provide better hemodynamic performance at rest and during exercise than stented valves, and are advocated in this group of patients. A new-generation stented valve, the Trifecta (St. Jude), has recently become available with improved hemodynamics. The aim of this study was to compare the hemodynamic performance of Freestyle (Medtronic) and Trifecta at rest and during exercise in patients with a small aortic annulus.
From September 2012 to September 2014, 22 patients with a native aortic annulus ≤ 23 mm underwent ergometric stress testing one year after aortic valve replacement with either a Trifecta (12 patients) or a Freestyle (10 patients) bioprosthesis as part of a randomized study.
The mean gradient at rest was 6.0 ± 2.3 mmHg for Trifecta and 4.3 ± 3.5 for Freestyle (p = 0.213). The mean gradient at peak of exercise was 9.7 ± 3.4 mmHg for Trifecta and 7.4 ± 5 mmHg for Freestyle (p = 0.243). No significant differences were found between the two prostheses regarding other hemodynamic parameters: effective orifice area, velocity index, and performance indexes.
Both the stented Trifecta and stentless Freestyle prostheses provide excellent hemodynamic results during physical stress in patients with a small aortic annulus. Our study confirms that Trifecta implantation results in low gradients at rest and during exercise and that the performance of Trifecta is similar to that of a stentless valve.
主动脉瓣环较小(即≤23mm)的患者给外科医生带来了挑战,因为他们发生患者-人工瓣膜不匹配的风险很高。无支架瓣膜在静息和运动时比有支架瓣膜具有更好的血流动力学性能,因此在这组患者中得到提倡。新一代有支架瓣膜Trifecta(圣犹达公司)最近已上市,其血流动力学有所改善。本研究的目的是比较Freestyle(美敦力公司)和Trifecta在静息和运动时,在主动脉瓣环较小的患者中的血流动力学性能。
作为一项随机研究的一部分,2012年9月至2014年9月期间,22例主动脉瓣环≤23mm的患者在接受主动脉瓣置换术后一年,接受了运动负荷试验,其中12例患者植入Trifecta生物瓣膜,10例患者植入Freestyle生物瓣膜。
Trifecta组静息时平均跨瓣压差为6.0±2.3mmHg,Freestyle组为4.3±3.5mmHg(p=0.213)。Trifecta组运动高峰时平均跨瓣压差为9.7±3.4mmHg,Freestyle组为7.4±5mmHg(p=0.243)。在其他血流动力学参数方面,两种人工瓣膜之间未发现显著差异:有效瓣口面积、速度指数和性能指数。
有支架的Trifecta和无支架的Freestyle人工瓣膜在主动脉瓣环较小的患者进行体力应激时均能提供优异的血流动力学结果。我们的研究证实,植入Trifecta后在静息和运动时跨瓣压差较低,且Trifecta的性能与无支架瓣膜相似。