Department of Radiology, University of Minnesota, 420 Delaware St. SE, MMC 292, Minneapolis, MN 55455; Department of Neurosurgery, University of Minnesota, 420 Delaware St. SE, MMC 292, Minneapolis, MN 55455.
Department of Radiology, University of Minnesota, 420 Delaware St. SE, MMC 292, Minneapolis, MN 55455.
J Vasc Interv Radiol. 2014 Apr;25(4):587-92. doi: 10.1016/j.jvir.2013.11.013.
Balloon-assisted embolization performed by delivering Onyx ethylene vinyl alcohol copolymer through a dual-lumen coaxial balloon microcatheter is a new technique for the management of peripheral vascular lesions. This technique does not require an initial reflux of Onyx to form around the tip of the microcatheter before antegrade flow of Onyx can commence. In a series of four patients who were treated with the use of this technique, the absence of significant reflux of Onyx was noted, as were excellent navigability and easy retrieval of the balloon microcatheter. However, in one patient, there was inadvertent adverse embolization of a digital artery, which was not caused by reflux of Onyx but could still be related to balloon inflation.
球囊辅助栓塞术通过双腔同轴微导管输送乙烯基醇共聚物 Onyx 来实现,是一种治疗外周血管病变的新技术。与传统技术不同,该技术不需要在开始顺行输送 Onyx 之前让 Onyx 回流到微导管尖端形成环。在使用该技术治疗的 4 名患者中,均未出现明显的 Onyx 回流,微导管具有良好的可操作性,且易于回收。然而,在 1 名患者中,意外发生了指动脉栓塞,这不是由 Onyx 回流引起的,而可能与球囊充气有关。