Suppr超能文献

分钟通气量作为频率适应性起搏参数的可靠性。

Reliability of minute ventilation as a parameter for rate responsive pacing.

作者信息

Lau C P, Antoniou A, Ward D E, Camm A J

机构信息

Department of Cardiological Sciences, St. George's Hospital Medical School, London.

出版信息

Pacing Clin Electrophysiol. 1989 Feb;12(2):321-30. doi: 10.1111/j.1540-8159.1989.tb02665.x.

Abstract

A minute ventilation sensing rate responsive pacemaker was implanted in 11 patients with bradycardias. Their mean age was 59 +/- 4 years (mean +/- SEM). The pacemaker measures minute ventilation by sensing intravascular impedance using a standard bipolar electrode. The rate responsive programming was simple: apart from ascribing an upper and lower rate, the only programmable parameter was the slope of rate response. This could be derived approximately by assessing the suggested slope value during an exercise test in the 'adaptive VVI' mode. Compared with exercise in the VVI mode, symptom limited treadmill tests in the rate responsive mode showed a 33% improvement of exercise capacity and a 44% improvement of cardiac output as determined noninvasively by continuous wave Doppler measurements of the ascending aortic blood flow. The pacing rate was appropriately increased during a variety of daily activities such as walking at different speeds and gradients, and ascending and descending stairs. Voluntary interference of the respiratory pattern such as during coughing and hyperventilation increased the pacing rate from a resting rate of 70 bpm to 111 +/- 10 and 86 +/- 4 bpm respectively. Continuous talking during exercise attenuated the expected rate response. The pacemaker can sense activity induced by arm swinging. In conclusion, the Meta pacemaker improved cardiac output and exercise capacity in patients with bradycardias. Its rate response was related to workload. Although voluntary interference affected the pacing rate, excessive rate acceleration was not encountered.

摘要

11例心动过缓患者植入了分钟通气量感知频率应答型起搏器。他们的平均年龄为59±4岁(均值±标准误)。该起搏器通过使用标准双极电极感知血管内阻抗来测量分钟通气量。频率应答编程很简单:除了设定上限和下限频率外,唯一可程控的参数是频率应答斜率。这可以通过在“自适应VVI”模式下的运动试验中评估建议的斜率值大致得出。与VVI模式下的运动相比,频率应答模式下症状受限的平板试验显示运动能力提高了33%,心输出量提高了44%,这是通过连续波多普勒测量升主动脉血流无创确定的。在各种日常活动中,如以不同速度和坡度行走、上下楼梯时,起搏频率会适当增加。呼吸模式的自主干扰,如咳嗽和过度通气时,起搏频率分别从静息时的70次/分增加到111±10次/分和86±4次/分。运动时持续说话会减弱预期的频率应答。该起搏器能感知手臂摆动引起的活动。总之,Meta起搏器改善了心动过缓患者的心输出量和运动能力。其频率应答与工作量有关。虽然自主干扰会影响起搏频率,但未出现过度的频率加速。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验