使用具有高时间分辨率的近红外光谱技术监测颈动脉支架置入术中滤网通畅情况。

Monitoring of filter patency during carotid artery stenting using near-infrared spectroscopy with high time-resolution.

机构信息

Division of Neurosurgery, Department of Neurological Surgery, Nihon University School of Medicine, 30-1 Oyaguchi-kamimachi, Itabashi-ku, Tokyo, 173-8610, Japan.

Department of Electrical and Electronics Engineering, Nihon University College of Engineering, NEWCAT Institute, Fukushima, Japan.

出版信息

Adv Exp Med Biol. 2014;812:325-331. doi: 10.1007/978-1-4939-0620-8_43.

Abstract

We aimed to evaluate the usefulness of a newly developed, near-infrared spectroscopy (NIRS) device for monitoring hemodynamic changes during carotid artery stenting (CAS), as a means to detect filter obstruction due to distal embolism. We evaluated 16 patients with internal carotid artery (ICA) stenosis during the CAS procedure, using a NIRS system that can monitor not only changes in oxygenation of hemoglobin (Hb), but also the fluctuation of oxyhemoglobin (oxy-Hb) synchronized with heartbeat. The NIRS system detected a marked decrease of oxy-Hb and an increase of deoxyhemoglobin (deoxy-Hb) during ICA occlusion in patients without anterior cross circulation (ACC). Patients with ACC showed much smaller changes. The analysis of oxy-Hb fluctuation made it possible to detect occurrence of no-flow in the absence of Hb concentration changes. The amplitude of oxy-Hb fluctuation in the no/slow-flow group was significantly smaller than that in the normal-flow group. Our results indicate that the present high time-resolution NIRS device, which can measure oxy-Hb fluctuation, is superior to conventional NIRS for detecting filter obstruction.

摘要

我们旨在评估一种新开发的近红外光谱(NIRS)设备在监测颈动脉支架置入术(CAS)期间血流动力学变化方面的有用性,作为检测因远端栓塞引起的滤器阻塞的一种手段。我们评估了 16 例颈内动脉(ICA)狭窄患者在 CAS 过程中使用 NIRS 系统的情况,该系统不仅可以监测血红蛋白(Hb)氧合变化,还可以监测与心跳同步的氧合血红蛋白(oxy-Hb)波动。在没有前向交叉循环(ACC)的患者中,NIRS 系统在 ICA 闭塞期间检测到 oxy-Hb 的明显下降和脱氧血红蛋白(deoxy-Hb)的增加。有 ACC 的患者变化较小。oxy-Hb 波动分析使得即使在 Hb 浓度变化不明显的情况下也能够检测到无血流。无/慢血流组的 oxy-Hb 波动幅度明显小于正常血流组。我们的结果表明,目前具有高时间分辨率的 NIRS 设备,能够测量 oxy-Hb 波动,在检测滤器阻塞方面优于传统 NIRS。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验