Arakawa M, Kambara K, Ito H, Hirakawa S, Umeda S, Hirose H
Second Department of Internal Medicine, Gifu University School of Medicine, Japan.
Pacing Clin Electrophysiol. 1989 Aug;12(8):1312-6. doi: 10.1111/j.1540-8159.1989.tb05044.x.
A 49-year-old male patient developed sensing failure (oversensing) 6 months after the implantation of a temperature sensing rate responsive pacemaker by the subclavian venipuncture method. Intermittent oversensing appeared in the sitting position, but did not appear in the supine position. Temperature telemetry showed an excessive fluctuation of the temperature data points while sitting and while doing a treadmill exercise test. Internal insulation damage was found approximately 31 cm from the distal tip of the explanted lead. The electrical resistance between one thermistor coil and the pacing coil changed from 9 kiloohms to 40 ohms when moderate pressure was applied to the outside lead in the fault area. This electrical shunt resulted from internal insulation damage that resulted from compression of the pacemaker lead between the first rib and the clavicle.
一名49岁男性患者在通过锁骨下静脉穿刺法植入温度感应频率应答式起搏器6个月后出现感知故障(过感知)。间歇性过感知出现在坐位时,而仰卧位时未出现。温度遥测显示在坐位和进行跑步机运动试验时温度数据点波动过大。在取出的导线远端尖端约31厘米处发现内部绝缘损坏。当在故障区域对外侧导线施加适度压力时,一个热敏电阻线圈与起搏线圈之间的电阻从9千欧变为40欧。这种电分流是由于起搏器导线在第一肋骨和锁骨之间受压导致内部绝缘损坏引起的。