永久性心脏起搏的电极和导线设计差异是否会转化为临床上可证实的差异?(烧结铂电极与活性玻璃和多孔碳电极的比较)

Do electrode and lead design differences for permanent cardiac pacing translate into clinically demonstrable differences? (Comparison of sintered platinum and activated vitreous and porous carbon electrodes).

作者信息

Bourke J P, Howell L, Murray A, Hill W E, Cowan J C, Beatt K, Errington J, Jameson S, Gold R G

机构信息

University Department of Cardiology, Freeman Hospital, Newcastle-Upon-Tyne, U.K.

出版信息

Pacing Clin Electrophysiol. 1989 Aug;12(8):1419-25. doi: 10.1111/j.1540-8159.1989.tb05057.x.

Abstract

A randomized prospective study was undertaken to compare the electrical performances of three permanent, endocardial, tined pacing leads with different electrode designs--sintered platinum, vitreous carbon, and porous carbon. Ninety-nine patients received one of the leads (S80 31; 423S 32; S100 36). Acute R wave amplitude and ST elevation of the native endocardial electrogram, voltage threshold, impedance, and current flow at four pulse durations (0.25-1.0 msec) were measured. Voltage thresholds were measured noninvasively at each of four pulse durations at 2 days and 1, 3, and 6 months after implantation. No significant differences were found in sensing properties, or current flow at threshold at 0.5 msec pulse duration. The 423S lead had a significantly higher impedance at threshold and both a higher impedance and lower current flow at 5 V. No significant differences in threshold voltages were found between the three leads at any pulse duration, at any of the assessed times after implantation. Six-month thresholds for the S80, 423S, and S100 leads were 1.18 +/- 0.35, 1.17 +/- 0.29, and 1.06 +/- 0.38 V respectively at 0.5 msec pulse duration. Differences between 'high performance' pacing leads need to be of a greater order of magnitude before they can be exploited to give any real clinical advantage to patients.

摘要

开展了一项随机前瞻性研究,以比较三种具有不同电极设计的永久性心内膜带倒刺起搏导线的电性能,这三种电极设计分别为烧结铂、玻璃碳和多孔碳。99名患者接受了其中一种导线(S80型31例;423S型32例;S100型36例)。测量了自身心内膜电图的急性R波振幅和ST段抬高、电压阈值、阻抗以及四个脉冲持续时间(0.25 - 1.0毫秒)下的电流。在植入后2天以及1、3和6个月时,在四个脉冲持续时间的每一个时间点非侵入性测量电压阈值。在感知特性或0.5毫秒脉冲持续时间的阈值电流方面未发现显著差异。423S导线在阈值时的阻抗显著更高,在5V时阻抗更高且电流更低。在植入后的任何评估时间,在任何脉冲持续时间下,三种导线之间的阈值电压均未发现显著差异。在0.5毫秒脉冲持续时间时,S80、423S和S100导线的6个月阈值分别为1.18±0.35、1.17±0.29和1.06±0.38V。“高性能”起搏导线之间的差异需要达到更大的数量级,才能被利用来为患者带来任何实际的临床优势。

相似文献

7

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验