Goicolea A, Chicote R, Adoue P, Gómez P, García-Cosío F
Servicio de Cardiología, Hospital Central de la Cruz Roja, Madrid.
Rev Esp Cardiol. 1990 Feb;43(2):80-3.
Low-output pacemaker programming depends on the absolute value of the chronic pacing threshold. To assess the chronic threshold with three different types of leads we have used Siemens-Vario pacemakers which allow us to measure the chronic voltage threshold noninvasively. Forty six patients had a 12 mm2 platinum tip lead; 38, 12 mm2 carbon tip lead an 15, 9 mm2 carbon tip lead (Siemens, 414-415, 412 and 423). Platinum leads longevity was 45 +/- 15.9 months; 12 mm2 carbon tip leads, 28 +/- 6.5 and 9 mm2 carbon tip leads, 17 +/- 4.5. Forty two months postimplantation chronic threshold (measured at 0.5 msec pulse-width) was 1.6 +/- 0.6 V for platinum electrodes and 1 +/- 0.3 V for 12 mm2 carbon tip leads (p less than 0.001). Twenty four months postimplantation 9 mm2 carbon electrodes chronic threshold was 1 +/- 0.4 V, and 1.7 +/- 0.5 V for platinum electrodes (p less than 0.001). There were no differences between carbon tip leads with 9 and 12 mm2 pacing area. The new activated carbon tip leads have lower chronic pacing thresholds than platinum electrodes with the same pacing area. Pacemaker-patients with these new type of leads can benefit of reduced output programming and increase their pacemaker longevity.
低输出起搏器程控取决于慢性起搏阈值的绝对值。为了用三种不同类型的电极导线评估慢性阈值,我们使用了西门子 - 瓦里欧起搏器,它使我们能够无创地测量慢性电压阈值。46例患者使用12平方毫米铂铱合金头电极导线;38例使用12平方毫米碳头电极导线,15例使用9平方毫米碳头电极导线(西门子,型号分别为414 - 415、412和423)。铂铱合金头电极导线的使用寿命为45±15.9个月;12平方毫米碳头电极导线为28±6.5个月,9平方毫米碳头电极导线为17±4.5个月。植入后42个月时,铂电极的慢性阈值(在脉宽0.5毫秒时测量)为1.6±0.6伏,12平方毫米碳头电极导线为1±0.3伏(p<0.001)。植入后24个月时,9平方毫米碳电极的慢性阈值为1±0.4伏,铂电极则为1.7±0.5伏(p<0.001)。9平方毫米和12平方毫米起搏面积的碳头电极导线之间无差异。新型活性炭头电极导线的慢性起搏阈值低于相同起搏面积的铂电极。使用这些新型电极导线的起搏器患者可受益于更低的输出程控,并延长起搏器使用寿命。