Murray A, Jameson S, Gold R G
Regional Medical Physics Department, Freeman, Hospital, Newcastle upon Tyne, United Kingdom.
Pacing Clin Electrophysiol. 1995 Dec;18(12 Pt 1):2150-4. doi: 10.1111/j.1540-8159.1995.tb04640.x.
A double-blind randomized study of three pacing leads, identical in all ways except for the surface of their electrodes, is presented. The leads were implanted in 30 patients. Ten patients (Group A) received a standard Laserdish lead (dish electrode with laser pores), 10 patients (Group B) received a platinum-iridium coated modified Laserdish electrode, and 10 patients (Group C) received a platinum-iridium coated dish electrode identical to that in Group B except for the absence of laser pores. All leads were implanted by the same operator via the subclavian vein and all patients received an Optima MPT pulse generator. Direct measurement at implantation of pacing threshold and endocardial electrograms showed no significant difference between the three groups (mean +/- standard deviation voltage thresholds 0.30 +/- 0.06 V at 0.5 ms and 0.21 +/- 0.05 V at 1.0-ms pulse durations, R wave amplitude 7.6 +/- 3.2 mV). Significant differences (P = 0.001) were found in threshold impedances at implant (Group A 940 omega, Group B 782 omega, Group C 673 omega). Ten further measurements of voltage threshold were made over the next 2 years. Overall, at a pulse duration of 1.0 ms, a significantly reduced (P < 0.025) threshold voltage was found for both the electrode with pores and platinum-iridium coating compared with pores alone (20%), and with pores and platinum-iridium coating compared with the coating alone (18%). There was no significant difference between using platinum-iridium alone or pores alone. We have shown that a controlled randomized study, using electrodes that are identical except for the characteristics being assessed, can enable effects of small differences in electrode design to be quantified.
本文介绍了一项双盲随机研究,该研究涉及三种起搏导线,除电极表面外其他方面均相同。这些导线被植入30名患者体内。10名患者(A组)接受标准的激光孔盘状导线(带有激光孔的盘状电极),10名患者(B组)接受铂铱涂层改良激光孔盘状电极,10名患者(C组)接受与B组相同的铂铱涂层盘状电极,但没有激光孔。所有导线均由同一名操作者经锁骨下静脉植入,所有患者均接受Optima MPT脉冲发生器。植入时起搏阈值和心内膜电图的直接测量结果显示,三组之间无显著差异(平均±标准差电压阈值:0.5 ms时为0.30±0.06 V,1.0 ms脉冲持续时间时为0.21±0.05 V,R波振幅为7.6±3.2 mV)。植入时的阈值阻抗存在显著差异(P = 0.001)(A组940Ω,B组782Ω,C组673Ω)。在接下来的2年中又进行了10次电压阈值测量。总体而言,在1.0 ms的脉冲持续时间下,与单独有孔电极相比,有孔且有铂铱涂层的电极阈值电压显著降低(P < 0.025)(降低20%),与单独有涂层相比,有孔且有铂铱涂层的电极阈值电压也显著降低(降低18%)。单独使用铂铱或单独使用孔之间无显著差异。我们已经表明,使用除被评估特征外其他均相同的电极进行对照随机研究,可以量化电极设计中微小差异的影响。