Department of Pathology, Immunology and Laboratory Medicine, University of Florida, Gainesville, FL 32610.
Division of Infectious Diseases, Department of Medicine, University of Florida, Gainesville, FL 32610.
Diagn Microbiol Infect Dis. 2014 Jul;79(3):293-7. doi: 10.1016/j.diagmicrobio.2014.03.014. Epub 2014 Mar 22.
We evaluated the accuracy and performance of the FilmArray Direct from Positive Blood Culture system (BCID) (BioFire Diagnostics, Salt Lake City, UT, USA) and the VITEK Mass Spectrometry System (Vitek MS; bioMerieux, Durham, NC, USA) to identify bacterial isolates from 161 positive blood culture bottles. The BCID uses multiplex PCR to identify 90-95% of common isolates to the genus or species/complex level as well as mecA, Van A/B, and bla(KPC) genes in approximately 1 hour. Of 151 monomicrobic isolates, the FilmArray correctly identified 48/49 (98%) to the genus and 84/84 (100%) to the species/complex level, while 18/151 (12%) gave no identification, as expected from the database. Mass spectrometry correctly identified 142/151 (94%) monomicrobic cultures to the genus level, 137/151 (91%) to the species level, with only 8/151(5%) giving no identification. Although mass spectrometry has a much larger database, the filtration system was cumbersome in contrast to the 3-5 minutes hands-on-time for the BCID.
我们评估了 FilmArray Direct 从阳性血培养系统 (BCID)(BioFire Diagnostics,盐湖城,美国)和 VITEK 质谱系统 (Vitek MS; bioMerieux,达勒姆,NC,美国)鉴定 161 个阳性血培养瓶中细菌分离物的准确性和性能。BCID 使用多重 PCR 鉴定 90-95%的常见分离物到属或种/复合水平,以及 mecA、Van A/B 和 bla(KPC)基因,大约需要 1 小时。在 151 个单微生物分离物中,FilmArray 正确鉴定了 48/49(98%)至属水平和 84/84(100%)至种/复合水平,而 18/151(12%)未鉴定,这与数据库中的预期相符。质谱正确鉴定了 142/151(94%)的单微生物培养物到属水平,137/151(91%)到种水平,只有 8/151(5%)未鉴定。尽管质谱具有更大的数据库,但与 BCID 的 3-5 分钟手动时间相比,过滤系统非常繁琐。