Routine clinical practice involves the application of diverse scanning parameters that can affect apparent diffusion coefficient (ADC) values. We evaluated interimager variability in ADC values with respect to their potential effect in clinical applications.
In 7 healthy volunteers, we obtained diffusion-weighted (DW) images using routine clinical parameters and 1.5- (n = 9) and 3-tesla (n = 3) magnetic resonance (MR) imagers from 5 different vendors, performing 84 MR imaging studies. To evaluate the differences in ADC values among the imagers, vendors, and magnetic field strengths, we measured the mean pixel values of the frontal white matter and thalamus (gray matter) in both cerebral hemispheres of the 7 volunteers and used repeated-measures analysis of variance for multiple comparisons.
The laterality of ADC values in the bilateral structures ranged from one to 3% for the 12 imagers. Although the relative difference in ADC values of white matter was 7% for scanners yielding the highest and lowest mean ADC values (P < 0.01), it was within 2 to 4% for instruments from the same vendors. For gray matter, the interimager difference was 4 to 12%, even among the same vendors (P < 0.05). Among the 3T imagers, the difference for white and gray matter was approximately 3%.
There were significant interimager differences in ADC values, especially with respect to gray matter. Taking into consideration the existing laterality, however, the differences among our 3T imagers may be acceptable despite the use of diverse scanning parameters. In routine clinical practice, the existing variability must be considered imager by imager.
常规临床实践涉及多种扫描参数的应用,这些参数可能会影响表观扩散系数(ADC)值。我们评估了不同成像仪之间ADC值的变异性及其在临床应用中的潜在影响。
在7名健康志愿者中,我们使用常规临床参数以及来自5家不同供应商的1.5特斯拉(n = 9)和3特斯拉(n = 3)磁共振(MR)成像仪获取扩散加权(DW)图像,共进行了84次MR成像研究。为了评估成像仪、供应商和磁场强度之间ADC值的差异,我们测量了7名志愿者双侧大脑半球额叶白质和丘脑(灰质)的平均像素值,并使用重复测量方差分析进行多重比较。
12台成像仪测得的双侧结构ADC值的左右差异范围为1%至3%。虽然产生最高和最低平均ADC值的扫描仪之间白质ADC值的相对差异为7%(P < 0.01),但同一供应商的仪器之间差异在2%至4%以内。对于灰质,即使在同一供应商的仪器之间,成像仪间差异也为4%至12%(P < 0.05)。在3T成像仪中,白质和灰质的差异约为3%。
ADC值在成像仪之间存在显著差异,尤其是在灰质方面。然而,考虑到现有的左右差异,尽管使用了不同的扫描参数,我们的3T成像仪之间的差异可能是可以接受的。在常规临床实践中,必须逐台成像仪考虑现有的变异性。