Yamauchi-Kawaura C, Fujii K, Akahane K, Yamauchi M, Narai K, Aoyama T, Katsu T, Obara S, Imai K, Ikeda M
Department of Radiological Sciences, Graduate School of Medicine, Nagoya University, 1-1-20 Daikominami, Higashi-ku, Nagoya 461-8673, Japan
Department of Radiological Sciences, Graduate School of Medicine, Nagoya University, 1-1-20 Daikominami, Higashi-ku, Nagoya 461-8673, Japan.
Radiat Prot Dosimetry. 2015 Feb;163(2):188-201. doi: 10.1093/rpd/ncu155. Epub 2014 May 12.
In this study, the authors developed age-specific physical head phantoms simulating the physique of Japanese children for dose evaluation in paediatric head computed tomography (CT) examinations. Anatomical structures at 99 places in 0-, 0.5-, 1- and 3-y-old Japanese patients were measured using DICOM viewer software from CT images, and the head phantom of each age was designed. For trial manufacture, a 3-y-old head phantom consisting of acrylic resin and gypsum was produced by machine processing. Radiation doses for the head phantom were measured with radiophotoluminescence glass dosemeters and Si-pin photodiode dosemeters. To investigate whether the phantom shape was suitable for dose evaluation, organ doses in the same scan protocol were compared between the 3-y-old head and commercially available anthropomorphic phantoms having approximately the same head size. The doses of organs in both phantoms were equivalent. The authors' designed paediatric head phantom will be useful for dose evaluation in paediatric head CT examinations.
在本研究中,作者开发了特定年龄的头部物理模型,以模拟日本儿童的体格,用于儿科头部计算机断层扫描(CT)检查中的剂量评估。使用DICOM查看器软件从CT图像测量0岁、0.5岁、1岁和3岁日本患者99个部位的解剖结构,并设计每个年龄的头部模型。为进行试制,通过机械加工制作了一个由丙烯酸树脂和石膏组成的3岁头部模型。使用放射光致发光玻璃剂量计和硅PIN光电二极管剂量计测量头部模型的辐射剂量。为研究模型形状是否适合剂量评估,比较了3岁头部模型与头部尺寸大致相同的市售人体模型在相同扫描方案下的器官剂量。两种模型的器官剂量相当。作者设计的儿科头部模型将有助于儿科头部CT检查中的剂量评估。