Abstract

Thromboelastography is used for assessment of coagulation and to guide administration of blood products peri-operatively. There is currently no method of standardisation in the UK, nor an approved method of proving quality. We investigated the reproducibility of thromboelastography by testing whole blood with no coagulation abnormality in three phases. Where a single operator performed multiple assays on the same blood sample at a single location, we found considerable variation, with 21% of R- and 25% of K-time measurements lying outside a set tolerance range (median ± 20%). Where samples were analysed by different operators in a single location, this finding was repeated. Where blood was transported in a citrated form for simultaneous analysis in multiple locations, results were more consistent, suggesting improved stability. Across all phases of testing there was good reproducibility of the maximum amplitude. Further examination of the results indicated less variation where analysis was performed on blood taken from the same kaolin vial compared with results from different vials. Our preliminary study indicates that R- and K-times may be highly variable, which we hypothesise may be due to variable mixing of blood and kaolin. We intend to repeat this study in the context of coagulopathy, where variability in results could potentially impact upon transfusion practice.

摘要

血栓弹力描记术用于评估凝血功能,并指导围手术期血液制品的使用。目前在英国没有标准化的方法,也没有经过批准的质量证明方法。我们通过在三个阶段测试无凝血异常的全血来研究血栓弹力描记术的可重复性。当单个操作者在单个地点对同一血样进行多次检测时,我们发现存在相当大的差异,21%的 R 时间和 25%的 K 时间测量值超出了设定的公差范围(中位数±20%)。当在同一地点由不同操作者分析样本时,会重复出现这种情况。当血液以柠檬酸盐形式运输并在多个地点同时进行分析时,结果更加一致,表明稳定性有所提高。在所有测试阶段,最大振幅的可重复性都很好。进一步分析结果表明,与来自不同小瓶的结果相比,从同一高岭土小瓶中抽取的血液进行分析时,变化较小。我们的初步研究表明,R 时间和 K 时间可能高度可变,我们假设这可能是由于血液和高岭土的混合不均匀所致。我们打算在凝血功能障碍的背景下重复这项研究,因为结果的可变性可能会对输血实践产生影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验