Tang X, Ens W, Mayer F, Standing K G, Westmore J B
Physics Department, University of Manitoba, Winnipeg, Canada.
Rapid Commun Mass Spectrom. 1989 Dec;3(12):443-8. doi: 10.1002/rcm.1290031210.
Daughter ions from decomposition of [M + H]+ parent ions have been observed in a time-of-flight mass spectrometer fitted with an ion mirror. Unit mass resolution is obtained for parent ions of masses up to several thousand u when the mirror voltage is set at a value determined by the usual velocity-focusing criterion for parent ions. Under this condition the daughter-ion resolution in the low-mass range suffers. Considerable improvement is obtained when the daughter-ion spectrum is examined in several segments, with the mirror voltage optimized for each segment. The daughter-ion spectrum from decay of metastable [M + H]+ parent ions in Substance P is measured using this technique. Reasons are suggested for differences between the spectrum obtained and a spectrum for the same compound recently obtained using a tandem double-focusing instrument.