Krintus Magdalena, Kozinski Marek, Boudry Pascal, Capell Nuria Estañ, Köller Ursula, Lackner Karl, Lefèvre Guillaume, Lennartz Lieselotte, Lotz Johannes, Herranz Antonio Mora, Nybo Mads, Plebani Mario, Sandberg Maria B, Schratzberger Wolfgang, Shih Jessie, Skadberg Øyvind, Chargui Ahmed Taoufik, Zaninotto Martina, Sypniewska Grazyna

出版信息

Clin Chem Lab Med. 2014 Nov;52(11):1657-65. doi: 10.1515/cclm-2014-0107.

Abstract

BACKGROUND

International recommendations highlight the superior value of cardiac troponins (cTns) for early diagnosis of myocardial infarction along with analytical requirements of improved precision and detectability. In this multicenter study, we investigated the analytical performance of a new high sensitive cardiac troponin I (hs-cTnI) assay and its 99th percentile upper reference limit (URL).

METHODS

Laboratories from nine European countries evaluated the ARCHITECT STAT high sensitive troponin I (hs-TnI) immunoassay on the ARCHITECT i2000SR/i1000SR immunoanalyzers. Imprecision, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ) linearity of dilution, interferences, sample type, method comparisons, and 99th percentile URLs were evaluated in this study.

RESULTS

Total imprecision of 3.3%-8.9%, 2.0%-3.5% and 1.5%-5.2% was determined for the low, medium and high controls, respectively. The lowest cTnI concentration corresponding to a total CV of 10% was 5.6 ng/L. Common interferences, sample dilution and carryover did not affect the hs-cTnI results. Slight, but statistically significant, differences with sample type were found. Concordance between the investigated hs-cTnI assay and contemporary cTnI assay at 99th percentile cut-off was found to be 95%. TnI was detectable in 75% and 57% of the apparently healthy population using the lower (1.1 ng/L) and upper (1.9 ng/L) limit of the LoD range provided by the ARCHITECT STAT hs-TnI package insert, respectively. The 99th percentile values were gender dependent.

CONCLUSIONS

The new ARCHITECT STAT hs-TnI assay with improved analytical features meets the criteria of high sensitive Tn test and will be a valuable diagnostic tool.

摘要

背景

国际指南强调心肌肌钙蛋白(cTn)在心肌梗死早期诊断中的卓越价值以及对提高精密度和可检测性的分析要求。在这项多中心研究中,我们调查了一种新型高敏心肌肌钙蛋白I(hs-cTnI)检测方法及其第99百分位上限参考值(URL)的分析性能。

方法

来自9个欧洲国家的实验室在ARCHITECT i2000SR/i1000SR免疫分析仪上评估了ARCHITECT STAT高敏肌钙蛋白I(hs-TnI)免疫检测方法。本研究评估了不精密度、空白限(LoB)、检测限(LoD)、定量限(LoQ)、稀释线性、干扰、样本类型、方法比较以及第99百分位URL。

结果

低、中、高对照品的总不精密度分别为3.3%-8.9%、2.0%-3.5%和1.5%-5.2%。总变异系数为10%时对应的最低cTnI浓度为5.6 ng/L。常见干扰、样本稀释和携带污染均未影响hs-cTnI检测结果。发现样本类型存在轻微但具有统计学意义的差异。在第99百分位临界值时,所研究的hs-cTnI检测方法与当代cTnI检测方法的一致性为95%。使用ARCHITECT STAT hs-TnI试剂盒说明书提供的较低(1.1 ng/L)和较高(1.9 ng/L)检测限范围,分别在75%和57%的表面健康人群中检测到TnI。第99百分位值存在性别差异。

结论

具有改进分析特性的新型ARCHITECT STAT hs-TnI检测方法符合高敏Tn检测标准,将成为一种有价值的诊断工具。