Rahimi Mahdi Salmani, Holmes James H, Wang Kang, Reeder Scott B, Korosec Frank R
Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Magn Reson Med. 2015 May;73(5):1926-31. doi: 10.1002/mrm.25315. Epub 2014 Jun 9.
Two-point fat-water separation methods are increasingly being used for chest and abdominal MRI and have recently been introduced for use in MR angiography of the lower extremities. With these methods, flowing spins can accumulate unintended phase shifts between the echo times. The purpose of this study is to demonstrate that these phase shifts can lead to inaccurate signals in the water and fat images.
In vitro experiments were conducted at 1.5T and 3.0T using a stenosis-mimicking phantom and a computer-controlled pump to image a range of physiologically relevant velocities.
In the phantom images acquired using bipolar readout gradients, fat-water signal inaccuracies were visible in regions of flow, with increasing severity as the flow rate was increased. Additionally, similar effects were observed in regions of high flow in clinical chest and liver exams. In the phantom images, the effect was eliminated by using a dual-pass method without bipolar readout gradients.
When using fat-water separation methods with bipolar readout gradients, phase shifts caused by the motion of spins can lead to signal inaccuracies in the fat and water images. These artifacts can be mitigated by using approaches that do not use bipolar readout gradients.
两点脂肪-水分离方法越来越多地用于胸部和腹部磁共振成像(MRI),最近已引入用于下肢磁共振血管造影。使用这些方法时,流动的自旋在回波时间之间可能会累积意外的相移。本研究的目的是证明这些相移会导致水和脂肪图像中的信号不准确。
在1.5T和3.0T场强下进行体外实验,使用模拟狭窄的体模和计算机控制的泵对一系列生理相关速度进行成像。
在使用双极读出梯度采集的体模图像中,流动区域可见脂肪-水信号不准确,且随着流速增加,严重程度增加。此外,在临床胸部和肝脏检查的高血流区域也观察到类似效应。在体模图像中,通过使用无双极读出梯度的双程方法消除了这种效应。
当使用具有双极读出梯度的脂肪-水分离方法时,自旋运动引起的相移会导致脂肪和水图像中的信号不准确。通过使用不使用双极读出梯度的方法可以减轻这些伪影。