Investigation on the reduction of electron contamination with a 6-MV x-ray beam.

作者信息

Parthasaradhi K, Prasad S G, Rao B M, Lee Y, Ruparel R, Garces R

机构信息

Division of Medical Physics, Evanston Hospital, Northwestern University Medical School, Illinois 60201.

出版信息

Med Phys. 1989 Jan-Feb;16(1):123-5. doi: 10.1118/1.596399.

Abstract

Experimental investigations have been carried out on the reduction of electron contamination of a 6-MV x-ray beam of Clinac model 1800 for square field sizes 5 X 5 to 30 X 30 cm2 in steps of 5 cm and for rectangular field sizes 19 X 7 and 7 X 19 cm2. The electron contamination of both the open beam and the beam with the tray can be effectively reduced by placing a lead foil filter immediately below the blocking tray. Measurements at 100-cm source-skin distance with filter in place showed a reduction in dose in the buildup region and also a displacement of the location of Dmax to greater depths, even for small field sizes such as 10 X 10 cm2.

摘要

针对Varian Clinac 1800型直线加速器6兆伏X射线束,进行了实验研究,研究范围包括方形野尺寸从5×5至30×30平方厘米(步长为5厘米)以及矩形野尺寸为19×7和7×19平方厘米的情况。通过在遮挡托盘正下方放置一块铅箔滤过器,可有效降低开放射野和带有托盘的射野中的电子污染。在源皮距100厘米处放置滤过器进行测量,结果显示,即使对于10×10平方厘米这样的小射野尺寸,建成区的剂量也有所降低,并且最大剂量点的位置向更深的深度偏移。