Pinto R S, Robbins E, Seidenwurm D
Department of Radiology, NYU Medical Center, NY 10016.
AJNR Am J Neuroradiol. 1989 Mar-Apr;10(2):407-10.
Utilizing the scanning electron microscope, we compared a new guidewire with copolymer coating with standard Teflon-coated, coiled-spring guidewires in both clinical and in vitro settings. Intense thrombogenicity was observed with the Teflon-coated guidewires with formed thrombi ranging in size from 50-100 microns. No formed thrombus was noted on any of the specimens of the copolymer guidewire, although isolated clumps of platelets and erythrocytes without fibrin strands were seen infrequently. We conclude that the copolymer guidewire is markedly less thrombogenic than Teflon-coated guidewires.
我们使用扫描电子显微镜,在临床和体外环境中,将一种带有共聚物涂层的新型导丝与标准的涂有特氟龙的螺旋弹簧导丝进行了比较。观察到涂有特氟龙的导丝具有强烈的血栓形成性,形成的血栓大小在50 - 100微米之间。共聚物导丝的任何标本上均未发现形成的血栓,尽管偶尔可见没有纤维蛋白丝的孤立血小板和红细胞团块。我们得出结论,共聚物导丝的血栓形成性明显低于涂有特氟龙的导丝。