Rosenkrantz Andrew B, Chandarana Hersh, Pfeuffer Josef, Triolo Michael J, Shaikh Mohammed Bilal, Mossa David J, Geppert Christian
Department of Radiology, NYU Langone Medical Center, 550 First Avenue, New York, NY, 10016, USA,
Abdom Imaging. 2015 Jan;40(1):120-6. doi: 10.1007/s00261-014-0181-2.
To assess impact of two-channel parallel transmission (pTx) with focused excitation [zoomed echo-planar imaging (EPI)] on image quality of prostate diffusion-weighted imaging (DWI) at 3T.
27 male volunteers (27 ± 8 years) underwent 3T prostate MRI using 2-channel radiofrequency-transmit system and 18-channel torso receive coil. Scans included EPI-DWI sequence (b values 50, 500, 1000 s/mm(2)) acquired both with standard sinc pulse and 2-channel pTX with focused excitation, each acquired at large-field-of-view (FOV) (20 × 20 cm) and small-FOV (14 × 14 cm). An abdominal radiologist scored b-1000 images and ADC maps for image quality measures. Sequences were compared using paired Wilcoxon tests.
pTx with focused excitation showed significant improvements compared with standard DWI on b-1000 images at large-FOV for the absence of wrap and overall image quality (p ≤ 0.049); on b-1000 images at small-FOV for reduced distortion of prostate, absence of ghosting, absence of wrap, clarity of prostate capsule, clarity of peripheral/transition zone boundary, clarity of peri-urethral region, and overall image quality (p ≤ 0.004); and on ADC maps at small-FOV for reduced distortion of prostate, sharpness of prostate, clarity of prostatic capsule, clarity of peri-urethral region, and overall image quality (p = 0.002-0.036). When compared with standard large-FOV images, small-FOV images obtained using pTx with focused excitation showed no significant difference on the b-1000 images for any feature (p ≥ 0.175), while showing significant improvements on the ADC maps in terms of reduced distortion, absence of ghosting, and absence of wrap (p = 0.010-0.030).
Zoomed DWI using 2-channel pTx reduced artifacts and improved image quality for 3T prostate DWI; benefit was most apparent for small-FOV images.
评估采用聚焦激发的两通道并行传输(pTx)[缩放回波平面成像(EPI)]对3T前列腺扩散加权成像(DWI)图像质量的影响。
27名男性志愿者(27±8岁)使用两通道射频发射系统和18通道体部接收线圈进行3T前列腺MRI检查。扫描包括EPI-DWI序列(b值为50、500、1000 s/mm²),分别采用标准 sinc 脉冲以及采用聚焦激发的两通道pTX进行采集,每种采集方式均在大视野(FOV)(20×20 cm)和小视野(14×14 cm)下进行。一名腹部放射科医生对b-1000图像和表观扩散系数(ADC)图进行图像质量评分。采用配对Wilcoxon检验对序列进行比较。
与标准DWI相比,采用聚焦激发的pTx在大视野下的b-1000图像上,在无包裹伪影和整体图像质量方面有显著改善(p≤0.049);在小视野下的b-1000图像上,在前列腺变形减少、无重影、无包裹伪影、前列腺包膜清晰度、外周/移行区边界清晰度、尿道周围区域清晰度以及整体图像质量方面有显著改善(p≤0.004);在小视野下的ADC图上,在前列腺变形减少、前列腺锐度、前列腺包膜清晰度、尿道周围区域清晰度以及整体图像质量方面有显著改善(p=0.002-0.036)。与标准大视野图像相比,采用聚焦激发的pTx获得的小视野图像在b-1000图像的任何特征上均无显著差异(p≥0.175),而在ADC图上,在变形减少、无重影和无包裹伪影方面有显著改善(p=0.010-0.030)。
采用两通道pTx的缩放DWI减少了伪影并改善了3T前列腺DWI的图像质量;对小视野图像的益处最为明显。