Jensen Maria M, Haase Christine, Zerahn Bo
Department of Clinical Physiology and Nuclear Medicine, Herlev Hospital, Herlev, Denmark.
Clin Physiol Funct Imaging. 2015 Nov;35(6):418-24. doi: 10.1111/cpf.12178. Epub 2014 Jun 30.
Estimation of left ventricular ejection fraction (LVEF) with 99MTc-HSA equilibrium radionuclide angiography (RA) is frequently used for assessing cardiac function. The purpose of this study was to investigate the interstudy repeatability of left (LV) and right (RV) ventricular volume and ejection fraction estimations, using a cadmium-zinc-telluride (CZT) SPECT camera.
Forty-six patients were scanned twice, interrupted by repositioning. Each acquisition was analysed twice by two experienced technologists. Interstudy and interobserver variations were calculated as the coefficient of variation (CV) and the 95% confidence interval (CI) for limits of agreement (LOA) between each sequence of analyses for each of the two acquisitions.
The Interstudy variations and 95% CI for LVEF, LV end systolic (LVESV) and end diastolic (LVEDV) volumes were 4.6% (-5.4 to 6.4), 9.3% (-6.90 to 5.20) and 7.0% (-13.9 to 11.1), respectively. For the right ventricle, the corresponding values were 11.9% (-9.40 to 10.8), 9.8% (-14.9 to 10.8) and 8.1% (-20.7 to 16.3).
The CZT detector camera has excellent reproducibility with regard to interstudy variation when assessing LV volumes and EF. Interstudy variation is considerably higher for RV volumes and EF, indicating a lack of consistency in tracing the boarders of the right ventricle.
采用99MTc-HSA平衡放射性核素血管造影(RA)估算左心室射血分数(LVEF)常用于评估心脏功能。本研究的目的是使用碲化镉锌(CZT)单光子发射计算机断层显像(SPECT)相机,研究左心室(LV)和右心室(RV)容积及射血分数估算的研究间重复性。
46例患者接受了两次扫描,中间有重新定位。每次采集由两名经验丰富的技术人员进行两次分析。研究间和观察者间差异计算为变异系数(CV)和两次采集中每次分析序列之间一致性界限(LOA)的95%置信区间(CI)。
LVEF、左心室收缩末期容积(LVESV)和舒张末期容积(LVEDV)的研究间差异及95%CI分别为4.6%(-5.4至6.4)、9.3%(-6.90至5.20)和7.0%(-13.9至11.1)。对于右心室,相应的值分别为11.9%(-9.40至10.8)、9.8%(-14.9至10.8)和8.1%(-20.7至16.3)。
在评估左心室容积和射血分数时,CZT探测器相机在研究间差异方面具有出色的可重复性。右心室容积和射血分数的研究间差异明显更高,表明在追踪右心室边界方面缺乏一致性。