Chen Yi-Chieh, Ko Chi-Lun, Yen Rouh-Fang, Lo Mei-Fang, Huang Yih-Hwen, Hsu Pei-Ying, Wu Yen-Wen, Cheng Mei-Fang
Department of Nuclear Medicine, National Taiwan University Hospital and National Taiwan University College of Medicine, No. 7, Chung-Shan South Road, Taipei, Taiwan.
Department of Nuclear Medicine, National Taiwan University Hospital Yun-Lin Branch, Douliou City, Yunlin County, Taiwan.
J Nucl Cardiol. 2016 Jun;23(3):348-61. doi: 10.1007/s12350-015-0367-1. Epub 2016 Mar 8.
We compared biventricular ejection fractions (EFs) from gated blood-pool single-photon emission computed tomography (SPECT) using a cadmium-zinc-telluride camera (CZT-SPECT) with planar equilibrium radionuclide angiography (ERNA) using a NaI gamma camera (NaI-planar). We also evaluated whether imaging time can be reduced without compromising image quality using the CZT camera.
Forty-eight patients underwent NaI-planar and CZT-SPECT on the same day. CZT-SPECT datasets were re-projected at an LAO orientation similar to ERNA acquisition, forming CZT-repro planar datasets. The resulting biventricular volumetric measurements and EFs were compared.
LVEF calculated from CZT-SPECT and CZT-repro correlated better with NaI-planar (r = 0.93 and 0.99, respectively) than RVEF (r = 0.76 and 0.82, respectively). Excellent intra-class correlation and low bias in intra-observer comparisons were observed for the biventricular EFs derived from three datasets. A wider limit of agreement in CZT-SPECT-derived LVEFs, lower correlation and significant bias for NaI-planar, and CZT-repro-derived RVEFs was found in the inter-observer analyses. Nonetheless, the imaging time can be reduced to 4 minutes without increasing variability in EFs using the CZT camera (P = NS).
LVEFs calculated from CZT-SPECT and CZT-repro correlated well with NaI-planar. CZT camera may reduce imaging time while preserving image quality in the assessment of biventricular EFs.
我们比较了使用碲化镉锌相机(CZT-SPECT)的门控血池单光子发射计算机断层扫描(SPECT)与使用碘化钠γ相机(NaI-平面)的平面平衡放射性核素血管造影(ERNA)所测得的双心室射血分数(EFs)。我们还评估了使用CZT相机在不影响图像质量的情况下是否可以缩短成像时间。
48例患者在同一天接受了NaI-平面和CZT-SPECT检查。将CZT-SPECT数据集以与ERNA采集相似的左前斜(LAO)方向重新投影,形成CZT-重新投影平面数据集。比较由此获得的双心室容积测量值和EFs。
与右心室射血分数(分别为r = 0.76和0.82)相比,由CZT-SPECT和CZT-重新投影计算得出的左心室射血分数与NaI-平面的相关性更好(分别为r = 0.93和0.99)。对于源自三个数据集的双心室EFs,在观察者内比较中观察到了极好的组内相关性和低偏差。在观察者间分析中,发现CZT-SPECT得出的左心室射血分数的一致性界限更宽,NaI-平面和CZT-重新投影得出的右心室射血分数的相关性较低且存在显著偏差。尽管如此,使用CZT相机可将成像时间缩短至4分钟,而不会增加EFs的变异性(P =无显著性差异)。
由CZT-SPECT和CZT-重新投影计算得出的左心室射血分数与NaI-平面相关性良好。在评估双心室EFs时,CZT相机可缩短成像时间,同时保持图像质量。