Otake Yosuke, Soutome Yoshihisa, Hirata Koji, Ochi Hisaaki, Bito Yoshitaka
Central Research Laboratory, Hitachi, Ltd.
Magn Reson Med Sci. 2014;13(3):199-205. doi: 10.2463/mrms.2013-0094. Epub 2014 Jul 2.
We developed a double-tuned radiofrequency (RF) coil using a novel circuit method to double tune for fluorine-19 (19F) and 1H magnetic resonance imaging, whose frequencies are very close to each other. The RF coil consists of 3 parallel-connected series inductor capacitor circuits. A computer simulation for our double-tuned RF coil with a phantom demonstrated that the coil has tuned resonant frequency and high sensitivity for both 19F and 1H. Drug distribution was visualized at 7 tesla using this RF coil and a rat administered perfluoro 15-crown-5-ether emulsion. The double-tune RF coil we developed may be a powerful tool for 19F and 1H imaging.