In the present study, bone samples from three patients treated in radiotherapy facilities in Poland were used for the determination of doses absorbed during radiotherapy. The samples were obtained during surgical treatments of patients performed due to medical indications. For the purpose of retrospective dosimetry, sensitivity of the radiation-induced EPR signal was individually calibrated in the samples by re-irradiation of the samples with known doses. The doses reconstructed in bones extracted within 6 months after irradiation were consistent with those calculated by treatment planning systems. The dose reconstructed in the bone removed 6 y after radiotherapy was ∼14% lower than the calculated one.