Accuracy of a digital impression system based on parallel confocal laser technology for implants with consideration of operator experience and implant angulation and depth.

作者信息

Giménez Beatriz, Özcan Mutlu, Martínez-Rus Francisco, Pradíes Guillermo

出版信息

Int J Oral Maxillofac Implants. 2014 Jul-Aug;29(4):853-62. doi: 10.11607/jomi.3343.

Abstract

PURPOSE

To evaluate the accuracy of a digital impression system based on parallel confocal red laser technology, taking into consideration clinical parameters such as operator experience and angulation and depth of implants.

MATERIALS AND METHODS

A maxillary master model with six implants (located bilaterally in the second molar, second premolar, and lateral incisor positions) was fitted with six polyether ether ketone scan bodies. One second premolar implant was placed with 30 degrees of mesial angulation; the opposite implant was positioned with 30 degrees of distal angulation. The lateral incisor implants were placed 2 or 4 mm subgingivally. Two experienced and two inexperienced operators performed intraoral scanning. Five different interimplant distances were then measured. The files obtained from the scans were imported with reverse-engineering software. Measurements were then made with a coordinate measurement machine, with values from the master model used as reference values. The deviations from the actual values were then calculated. The differences between experienced and inexperienced operators and the effects of different implant angulations and depths were compared statistically.

RESULTS

Overall, operator 3 obtained significantly less accurate results. The angulated implants did not significantly influence accuracy compared to the parallel implants. Differences were found in the amount of error in the different quadrants. The second scanned quadrant had significantly worse results than the first scanned quadrant. Impressions of the implants placed at the tissue level were less accurate than implants placed 2 and 4 mm subgingivally.

CONCLUSIONS

The operator affected the accuracy of measurements, but the performance of the operator was not necessarily dependent on experience. Angulated implants did not decrease the accuracy of the digital impression system tested. The scanned distance affected the predictability of the accuracy of the scanner, and the error increased with the increased length of the scanned section.

摘要

目的

考虑到诸如操作者经验、种植体角度和深度等临床参数,评估基于平行共焦红色激光技术的数字印模系统的准确性。

材料与方法

一个带有六个种植体(双侧位于第二磨牙、第二前磨牙和侧切牙位置)的上颌主模型安装了六个聚醚醚酮扫描体。一个第二前磨牙种植体以30度的近中角度植入;相对的种植体以30度的远中角度定位。侧切牙种植体植入龈下2或4毫米处。两名经验丰富的操作者和两名经验不足的操作者进行口内扫描。然后测量五个不同的种植体间距离。从扫描中获得的文件用逆向工程软件导入。然后用坐标测量机进行测量,将主模型的值用作参考值。然后计算与实际值的偏差。对经验丰富和经验不足的操作者之间的差异以及不同种植体角度和深度的影响进行统计学比较。

结果

总体而言,操作者3获得的结果准确性明显较低。与平行种植体相比,有角度的种植体对准确性没有显著影响。在不同象限的误差量上发现了差异。第二个扫描象限的结果明显比第一个扫描象限差。在组织水平植入的种植体印模不如植入龈下2毫米和4毫米处的种植体准确。

结论

操作者影响测量的准确性,但操作者的表现不一定取决于经验。有角度的种植体并没有降低所测试数字印模系统的准确性。扫描距离影响扫描仪准确性的可预测性,并且误差随着扫描段长度的增加而增加。