Suppr超能文献

Fracture of the cobalt-chromium modular femoral neck component in total hip arthroplasty.

作者信息

Mencière M-L, Amouyel T, Taviaux J, Bayle M, Laterza C, Mertl P

机构信息

Service de chirurgie orthopédique et traumatologique, CHU d'Amiens, hôpital Nord, 1, place Victor-Pauchet, 80054 Amiens, France.

Service de chirurgie orthopédique et traumatologique, CHU d'Amiens, hôpital Nord, 1, place Victor-Pauchet, 80054 Amiens, France.

出版信息

Orthop Traumatol Surg Res. 2014 Sep;100(5):565-8. doi: 10.1016/j.otsr.2014.03.027. Epub 2014 Aug 4.

Abstract

Total hip prostheses using cervico-metaphyseal modularity were designed to better replicate the geometry of the native extra-medullary femur. However, they are associated with numerous complications including corrosion, disassembly, pseudotumours and, most notably, fractures of the modular neck. All reported cases of modular neck fractures occurred with titanium components (Ti-6Al-4V). To prevent this weakness, manufacturers developed modular necks made of cobalt-chromium (Co-Cr). We report a fracture of a long, 8° varus, Co-Cr modular neck connected to a 36-mm short (-3.5mm) femoral head. The fracture occurred 22 months post-implantation in a woman who had a low level of physical activity and a body mass index of 28.7 kg/m(2). To our knowledge, this case is the first reported instance of Co-Cr modular neck fracture. It may challenge the wisdom of further developing this modularity design, as our patient had none of the known risk factors for modular neck fracture. In addition, cases of pseudotumour have been reported with Co-Cr modular necks subjected to fretting corrosion, which contributed to the fracture in our patient.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验