全髋关节置换翻修术后模块化股骨颈骨折:有限元分析

Modular femoral neck failure after revision of a total hip arthroplasty: a finite element analysis.

作者信息

Zajc Jan, Predan Jožef, Gubeljak Nenad, Moličnik Andrej, Fokter Samo K

机构信息

Faculty of Medicine, University of Maribor, Maribor, Slovenia.

Faculty of Mechanical Engineering, University of Maribor, Maribor, Slovenia.

出版信息

Eur J Orthop Surg Traumatol. 2019 Apr;29(3):717-723. doi: 10.1007/s00590-018-2314-8. Epub 2018 Oct 13.

Abstract

The authors report on a case of modular femoral neck fracture which appeared 21 months after revision of acetabular component. The revision surgery was performed 8 years after the primary total hip arthroplasty due to aseptic loosening of the acetabular component. During acetabular revision, the primary implanted short (S, - 3.5 mm) femoral head was also exchanged with extra-long (XL, + 7.0 mm) femoral head fitting the modular femoral neck with a longer lever arm. Numerical analysis has shown that this has resulted in a 19.9% increase in tensile stress at the neck-stem coupling during normal walking cycle. This could result in microcrack initiation and propagation and finally lead to modular neck failure of the otherwise well-fixed stem. Surgeons should avoid excessive loading of the exchangeable neck (dual-modular) femoral stem designs as the stem-neck couplings are subject to corrosion and are not as reliable as monoblock stems.

摘要

作者报告了一例模块化股骨颈骨折病例,该骨折出现在髋臼组件翻修术后21个月。翻修手术是在初次全髋关节置换术后8年因髋臼组件无菌性松动而进行的。在髋臼翻修期间,最初植入的短(S,-3.5毫米)股骨头也被换成了超长(XL,+7.0毫米)股骨头,该股骨头与模块化股骨颈适配,具有更长的杠杆臂。数值分析表明,这导致在正常步行周期中,颈-柄连接处拉应力增加了19.9%。这可能会导致微裂纹的萌生和扩展,最终导致原本固定良好的柄出现模块化颈部失效。外科医生应避免过度加载可更换颈部(双模块化)股骨柄设计,因为柄-颈连接处容易受到腐蚀,且不如整体式柄可靠。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验