Suppr超能文献

激光全瓷冠去除——一项实验室原理验证研究——第2阶段冠脱粘时间

Laser all-ceramic crown removal-a laboratory proof-of-principle study-phase 2 crown debonding time.

作者信息

Rechmann Peter, Buu Natalie C H, Rechmann Beate M T, Finzen Frederick C

机构信息

School of Dentistry, Department of Preventive and Restorative Dental Sciences, University of California at San Francisco, 707 Parnassus Avenue, San Francisco, California, 94143.

出版信息

Lasers Surg Med. 2014 Oct;46(8):636-43. doi: 10.1002/lsm.22280. Epub 2014 Aug 13.

Abstract

BACKGROUND AND OBJECTIVES

The removal of all-ceramic crowns is a time consuming procedure in the dental office. Little research has been done in alternative removal techniques for all-ceramic crowns. The objective of the second phase of this proof-of-principle laboratory pilot study was to evaluate whether Ivoclar Vivadent all-ceramic crowns can be efficiently removed from natural teeth without damage to the underlying tooth structure using an Erbium laser.

STUDY DESIGN/MATERIALS AND METHODS: The ceramic materials used were IPS E.max CAD Lithium-disilicate (LS2 ) (E.max CAD) and IPS E.max ZirCAD Zirconium-oxide (ZrO2 ) (ZirCAD) (Ivoclar, Vivadent, Liechtenstein). Molars, either as stand-alone teeth or placed in an artificial row of teeth, were prepared to receive all-ceramic crowns. Copings and full contour crowns with either featheredge or regular margins were produced. The all-ceramic crowns were bonded to the teeth with Ivoclar Multilink Automix. The time for Er:YAG laser debonding of each crown was then measured. The Er:YAG (LiteTouch, Syneron, Yokneam, Israel) was used with an 1,100 µm diameter fiber tip with energies up to 600 mJ per pulse (wavelength 2,940 nm, 10 Hz repetition rate, pulse duration 100 µs at 126 mJ/pulse, and 400 µs at 590 mJ/pulse). The irradiation was applied at a distance of 10 mm from the crown surface following a defined pattern. Air-water spray was applied to the crowns at a rate of 67 ml/minute.

RESULTS

All of the all-ceramic crowns were successfully debonded with the laser. On average, an all-ceramic E.max CAD crown was debonded in 190 ± 92 seconds (average ± SD). The debonding time for ZirCAD featheredge crowns was 226 ± 105 seconds and for ZirCAD crowns with regular margins it was 312 ± 102 seconds. No crowns fractured and no damage to the underlying dentin was detected. The bonding cement deteriorated due to the Er:YAG irradiation. Additionally, no carbonization at the dentin/cement interface was observed.

CONCLUSION

Er:YAG laser energy can successfully be used to efficiently debond all-ceramic full contour crowns from natural teeth without damage to the underlying tooth structure. Lasers Surg. Med. 46:636-643, 2014. © 2014 Wiley Periodicals, Inc.

摘要

背景与目的

在牙科诊所中,去除全瓷冠是一个耗时的过程。关于全瓷冠替代去除技术的研究较少。本原理验证实验室初步研究第二阶段的目的是评估是否可以使用铒激光从天然牙上有效去除义获嘉威瓦登特全瓷冠,同时不损伤其下方的牙体结构。

研究设计/材料与方法:所使用的陶瓷材料为IPS E.max CAD二硅酸锂(LS2)(E.max CAD)和IPS E.max ZirCAD氧化锆(ZrO2)(ZirCAD)(义获嘉,威瓦登特,列支敦士登)。制备磨牙,无论是单独的牙齿还是置于人造牙列中,以安装全瓷冠。制作了带有刃状边缘或常规边缘的基底冠和全冠。全瓷冠用义获嘉Multilink Automix粘结到牙齿上。然后测量每颗冠用铒钇铝石榴石激光脱粘的时间。使用的铒钇铝石榴石激光(LiteTouch,赛诺龙,以色列约克奈姆)配有直径1100μm的光纤头,每个脉冲能量高达600mJ(波长2940nm,重复频率10Hz,在126mJ/脉冲时脉冲持续时间100μs,在590mJ/脉冲时脉冲持续时间400μs)。按照规定的模式在距冠表面10mm的距离处进行照射。以67ml/分钟的速率向冠喷水雾。

结果

所有全瓷冠均成功用激光脱粘。平均而言,一颗全瓷E.max CAD冠脱粘用时190±92秒(平均值±标准差)。ZirCAD刃状边缘冠的脱粘时间为226±105秒,ZirCAD常规边缘冠的脱粘时间为312±102秒。没有冠发生破裂,也未检测到对下方牙本质的损伤。粘结剂因铒钇铝石榴石激光照射而变质。此外,在牙本质/粘结剂界面未观察到碳化现象。

结论

铒钇铝石榴石激光能量可成功用于从天然牙上有效脱粘全瓷全冠,且不损伤下方的牙体结构。《激光外科与医学》46卷:636 - 643页,2014年。©2014威利期刊公司

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验