Er:YAG laser is a non-destructive tool for debonding of laminate veneers. This study investigated the effect of different laser powers on the pulp temperature and the time required to perform the debonding of lithium disilicate laminate veneers with different thicknesses. The labial enamel of 48 maxillary central incisors was flattened and polished. The teeth were restored with flat lithium disilicate ceramic veneers (4.0 mm×6.0 mm) with one of two different thicknesses (0.5 and 1.0 mm). Veneer debonding was performed with an Er:YAG laser with a wavelength of 2940 nm, pulse duration of 100 μm (VSP mode), 10 Hz, and one of the three laser power settings: 1.5 W (150 mJ), 3.0 W (300 mJ). and 5.4 W (540 mJ) (n=8). Veneer detachment time and intra-pulp temperature change (ΔT) were measured. Statistical analysis was performed using the two-way ANOVA and Bonferroni's post hoc test (α=0.05). The correlation between debonding time and temperature change was calculated using Pearson's correlation. The longest time was recorded to remove the 1.0-mm veneer at 1.5 W (<0.05) and the shortest time was recorded when deboning the 0.5 mm veneer with 5.4 W (<0.05). ΔT decreased significantly with increasing laser power. A low correlation was found between time and ΔT (R=0.113). Laser power and veneer thickness are important factors for veneer debonding; thinner veneers are removed faster. When debonding thick veneers, 5.4W laser power is more efficient and causes fewer changes to the pulp temperature.
铒钇铝石榴石激光是一种用于去除层压贴面的无损工具。本研究调查了不同激光功率对牙髓温度的影响,以及去除不同厚度的二硅酸锂层压贴面所需的时间。将48颗上颌中切牙的唇面釉质磨平并抛光。用两种不同厚度(0.5毫米和1.0毫米)之一的扁平二硅酸锂陶瓷贴面(4.0毫米×6.0毫米)修复这些牙齿。使用波长为2940纳米、脉冲持续时间为100微秒(VSP模式)、频率为10赫兹的铒钇铝石榴石激光,以及三种激光功率设置之一:1.5瓦(150毫焦)、3.0瓦(300毫焦)和5.4瓦(540毫焦)(n = 8)进行贴面去除。测量贴面分离时间和牙髓内温度变化(ΔT)。使用双向方差分析和Bonferroni事后检验(α = 0.05)进行统计分析。使用Pearson相关性计算去除时间与温度变化之间的相关性。在1.5瓦时去除1.0毫米贴面的时间最长(<0.05),在5.4瓦去除0.5毫米贴面时记录到最短时间(<0.05)。随着激光功率的增加,ΔT显著降低。发现时间与ΔT之间的相关性较低(R = 0.113)。激光功率和贴面厚度是贴面去除的重要因素;较薄的贴面去除速度更快。去除厚贴面时,5.4瓦的激光功率更有效,对牙髓温度的影响更小。