Wu Aaron Yu-Jen, Hsu Jui-Ting, Huang Heng-Li
*Assistant Professor and Director, Department of Dentistry, Chang Gung Memorial Hospital, Chang Gung University College of Medicine, Kaohsiung, Taiwan. †Associate Professor, Biomechanics Res. Lab, School of Dentistry, China Medical University, Taichung, Taiwan. ‡Professor, Biomechanics Res. Lab, School of Dentistry, China Medical University, Taichung, Taiwan.
Implant Dent. 2014 Oct;23(5):534-8. doi: 10.1097/ID.0000000000000108.
The study compared the implant mobility and surrounding bone strain between the titanium-zirconium (Ti-Zr) alloy and the commercial pure (CP) Ti implants.
The mobility--quantified as the implant stability quotient (ISQ) and Periotest value (PTV)--of implants constructed from Ti-Zr alloy and CP Ti placed into artificial type-2 jawbone models were measured. Specimens were tested by applying 190 N vertically or at 30 degrees laterally. Peak values of the principal strains of bone were recorded by rosette strain gauges with a data acquisition system and were analyzed statistically using Wilcoxon rank-sum test.
PTV and ISQ values did not differ significantly between the Ti-Zr and CP Ti implants (P > 0.01). Under vertical loading, the peak bone strains did not differ significantly between the Ti-Zr and CP Ti specimens (P > 0.006). However, the peak strains were 52% lower around the Ti-Zr implant than around the Ti implant on the buccal side of bone under lateral loading (P < 0.001).
The implant material (Ti-Zr alloy vs CP Ti) had no effect on the mobility of small-diameter dental implants. However, using Ti-Zr alloy as an implant material decreased the periimplant bone strain under lateral loading in this pilot study.
本研究比较了钛锆(Ti-Zr)合金种植体和商业纯钛(CP Ti)种植体的植入物移动性及周围骨应变。
测量了由Ti-Zr合金和CP Ti制成的种植体植入人工2型颌骨模型后的移动性(以种植体稳定性商数(ISQ)和牙周探针值(PTV)量化)。通过垂直施加190 N或横向30度施加力对标本进行测试。用应变片式应变计和数据采集系统记录骨主应变的峰值,并采用Wilcoxon秩和检验进行统计学分析。
Ti-Zr种植体和CP Ti种植体的PTV和ISQ值无显著差异(P>0.01)。在垂直加载下,Ti-Zr标本和CP Ti标本的峰值骨应变无显著差异(P>0.006)。然而,在横向加载下,Ti-Zr种植体周围骨颊侧的峰值应变比Ti种植体周围低52%(P<0.001)。
种植体材料(Ti-Zr合金与CP Ti)对小直径牙种植体的移动性无影响。然而,在本初步研究中,使用Ti-Zr合金作为种植体材料可降低横向加载下种植体周围的骨应变。