使用锥形束计算机断层扫描图像评估头颈部调强放射治疗的治疗剂量分布:一项方法学研究。

Evaluating the therapeutic dose distribution of intensity-modulated radiation therapy for head and neck with cone-beam computed tomography image: a methodological study.

作者信息

Zhang Guang-shun, Huang Shao-min, Chen Cui, Xu Sen-kui, Zhang Dan-dan, Deng Xiao-wu

机构信息

Department of Radiation Oncology, Sun Yat-Sen University Cancer Center, State Key Laboratory of Oncology in Southern China, Collaborative Innovation Center for Cancer Medicine, 651 Dong Feng Road East, Guangzhou 510060, China.

出版信息

Biomed Res Int. 2014;2014:326532. doi: 10.1155/2014/326532. Epub 2014 Aug 17.

Abstract

An approximate correction method for the CT value-electron density curve of CBCT was established, through comparison and fitting with FBCT images, and applied to evaluate the therapeutic dose of IMRT. The precision of using CBCT for plan calculation was validated by comparing the dose distribution between CBCT- and FBCT-based IMRT plans. Also setup deviations were simulated to evaluate the ability of the CBCT-based calculation for detecting the dose errors caused by positioning deviation. The gamma comparison between CBCT- and FBCT-based dose computations showed that the pass rates of (2%, 2 mm) criteria were better than 97.60 ± 0.83% and 97.74 ± 2.08% in the phantom and 10 NPC cases. When setup deviation was introduced into CBCT-based dose calculation, the gamma pass rate significantly decreased while the volumetric doses of the targets and some normal organs exhibited different changes compared to the original plan. Our results validated the above CT value-electron density correction which reduced the difference between CBCT- and FBCT-based IMRT plan calculation for NPC to less than 2%. Online CBCT-based dose calculation can be used to reflect and evaluate the dose distribution discrepancy caused by setup deviation and structure changes during the treatment, ensuring more effective quality control of IMRT treatment.

摘要

通过与平板CT(FBCT)图像进行比较和拟合,建立了一种适用于锥束CT(CBCT)的CT值-电子密度曲线的近似校正方法,并将其应用于调强放射治疗(IMRT)治疗剂量的评估。通过比较基于CBCT和FBCT的IMRT计划之间的剂量分布,验证了使用CBCT进行计划计算的精度。同时模拟了摆位偏差,以评估基于CBCT的计算检测由定位偏差引起的剂量误差的能力。基于CBCT和FBCT的剂量计算之间的伽马比较表明,在体模和10例鼻咽癌(NPC)病例中,(2%,2毫米)标准的通过率分别优于97.60±0.83%和97.74±2.08%。当将摆位偏差引入基于CBCT的剂量计算时,伽马通过率显著降低,而靶区和一些正常器官的体积剂量与原计划相比呈现出不同的变化。我们的结果验证了上述CT值-电子密度校正,该校正将基于CBCT和FBCT的NPC的IMRT计划计算之间的差异降低到小于2%。基于在线CBCT的剂量计算可用于反映和评估治疗期间由摆位偏差和结构变化引起的剂量分布差异,确保IMRT治疗更有效的质量控制。