Validation of a MOSFET dosemeter system for determining the absorbed and effective radiation doses in diagnostic radiology.

作者信息

Manninen A-L, Kotiaho A, Nikkinen J, Nieminen M T

机构信息

Department of Diagnostic Radiology, Oulu University Hospital, PO Box 50, Oulu FI-90029, Finland Medical Research Center Oulu, Oulu University Hospital and University of Oulu, Oulu, Finland

Department of Diagnostic Radiology, Oulu University Hospital, PO Box 50, Oulu FI-90029, Finland Department of Physical Sciences, Biophysics, University of Oulu, PO Box 3000, Oulu FI-90014, Finland Medical Research Center Oulu, Oulu University Hospital and University of Oulu, Oulu, Finland.

出版信息

Radiat Prot Dosimetry. 2015 Apr;164(3):361-7. doi: 10.1093/rpd/ncu283. Epub 2014 Sep 11.

Abstract

This study aimed to validate a MOSFET dosemeter system for determining absorbed and effective doses (EDs) in the dose and energy range used in diagnostic radiology. Energy dependence, dose linearity and repeatability of the dosemeter were examined. The absorbed doses (ADs) were compared at anterior-posterior projection and the EDs were determined at posterior-anterior, anterior-posterior and lateral projections of thoracic imaging using an anthropomorphic phantom. The radiation exposures were made using digital radiography systems. This study revealed that the MOSFET system with high sensitivity bias supply set-up is sufficiently accurate for AD and ED determination. The dosemeter is recommended to be calibrated for energies <60 and >80 kVp. The entrance skin dose level should be at least 5 mGy to minimise the deviation of the individual dosemeter dose. For ED determination, dosemeters should be implanted perpendicular to the surface of the phantom to prevent the angular dependence error.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验