Takahashi Shigekiyo, Kawasaki Masanori, Miyata Shusaku, Suzuki Keita, Yamaura Makoto, Ido Takahisa, Aoyama Takuma, Fujiwara Hisayoshi, Minatoguchi Shinya
Department of Cardiology, Gifu University Gradual School of Medicine, 1-1 Yanagido, Gifu, 501-1194, Japan.
Department of Cardiology, Kizawa Memorial Hospital, Gifu, Japan.
Heart Vessels. 2016 Jan;31(1):29-37. doi: 10.1007/s00380-014-0577-5. Epub 2014 Sep 13.
Recently, a new generation of multi-detector row computed tomography (CT) with 320-detector rows (DR) has become available in the clinical settings. The purpose of the present study was to determine the cutoff values of Hounsfield unit (HU) for discrimination of plaque components by comparing HU of coronary plaques with integrated backscatter intravascular ultrasound (IB-IVUS) serving as a gold standard. Seventy-seven coronary atherosclerotic lesions in 77 patients with angina were visualized by both 320-DR CT (Aquilion One, Toshiba, Japan) and IB-IVUS at the same site. To determine the thresholds for discrimination of plaque components, we compared HU with IB values as a gold standard. Optimal thresholds were determined from receiver operating characteristic (ROC) curves analysis. The HU values of lipid pool (n = 115), fibrosis (n = 93), vessel lumen and calcification (n = 73) were 28 ± 19 HU (range -18 to 69 HU), 98 ± 31 HU (44 to 195 HU), 357 ± 65 HU (227 to 534 HU) and 998 ± 236 HU (366 to 1,489 HU), respectively. The thresholds of 56 HU, 210 HU and 490 HU were the most reliable predictors of lipid pool, fibrosis, vessel lumen and calcification, respectively. Lipid volume measured by 320-DR CT was correlated with that measured by IB-IVUS (r = 0.63, p < 0.05), whereas fibrous volume measured by 320-DR CT was not. Lipid volume measured by 320-DR CT was correlated with that measured by IB-IVUS, whereas fibrous volume was not correlated with that measured by IB-IVUS because manual exclusion of the outside of vessel hindered rigorous discrimination between fibrosis and extravascular components.
最近,新一代具有320排探测器(DR)的多排探测器计算机断层扫描(CT)已应用于临床。本研究的目的是通过将冠状动脉斑块的Hounsfield单位(HU)与作为金标准的集成背向散射血管内超声(IB-IVUS)进行比较,确定用于区分斑块成分的HU截止值。对77例心绞痛患者的77个冠状动脉粥样硬化病变部位同时进行了320排DR CT(日本东芝Aquilion One)和IB-IVUS检查。为了确定区分斑块成分的阈值,我们将HU与作为金标准的IB值进行比较。通过受试者操作特征(ROC)曲线分析确定最佳阈值。脂质池(n = 115)、纤维化(n = 93)、血管腔和钙化(n = 73)的HU值分别为28±19 HU(范围-18至69 HU)、98±31 HU(44至195 HU)、357±65 HU(227至534 HU)和998±236 HU(366至1489 HU)。56 HU、210 HU和490 HU的阈值分别是脂质池、纤维化、血管腔和钙化的最可靠预测指标。320排DR CT测量的脂质体积与IB-IVUS测量的脂质体积相关(r = 0.63,p < 0.05),而320排DR CT测量的纤维体积与IB-IVUS测量的纤维体积不相关。320排DR CT测量的脂质体积与IB-IVUS测量的脂质体积相关,而纤维体积与IB-IVUS测量的纤维体积不相关,因为手动排除血管外部阻碍了对纤维化和血管外成分的严格区分。