Chan Maria F, Cohen Gil'ad N, Deasy Joseph O
Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, NY, USA
Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, NY, USA.
Technol Cancer Res Treat. 2015 Jun;14(3):298-304. doi: 10.1177/1533034614547447. Epub 2014 Sep 16.
To evaluate visibility, artifacts, and distortions of various commercial markers in magnetic resonance imaging (MRI), computer tomography (CT), and ultrasound imaging used for radiotherapy planning and treatment guidance.
We compare 2 solid gold markers, 4 gold coils, and 1 polymer marker from 3 vendors. Imaging modalities used were 3-T and 1.5-T GE MRIs, Siemens Sequoia 512 Ultrasound, Phillips Big Bore CT, Varian Trilogy linear accelerator (cone-beam CT [CBCT], on-board imager kilovoltage [OBI-kV], electronic portal imaging device megavoltage [EPID-MV]), and Medtronic O-ARM CBCT. Markers were imaged in a 30 × 30 × 10 cm(3) custom bolus phantom. In one experiment, Surgilube was used around the markers to reduce air gaps. Images were saved in Digital Imaging and Communications in Medicine (DICOM) format and analyzed using an in-house software. Profiles across the markers were used for objective comparison of the markers' signals. The visibility and artifacts/distortions produced by each marker were assessed qualitatively and quantitatively.
All markers are visible in CT, CBCT, OBI-kV, and ultrasound. Gold markers below 0.75 mm in diameter are not visible in EPID-MV images. The larger the markers, the more CT and CBCT image artifacts there are, yet the degree of the artifact depends on scan parameters and the scanner itself. Visibility of gold coils of 0.75 mm diameter or larger is comparable across all imaging modalities studied. The polymer marker causes minimal artifacts in CT and CBCT but has poor visibility in EPID-MV. Gold coils of 0.5 mm exhibit poor visibility in MRI and EPID-MV due to their small size. Gold markers are more visible in 3-T T1 gradient-recalled echo than in 1.5-T T1 fast spin-echo, depending on the scan sequence. In this study, all markers are clearly visible on ultrasound.
All gold markers are visible in CT, CBCT, kV, and ultrasound; however, only the large diameter markers are visible in MV. When MR and EPID-MV imagers are used, the selection of fiducial markers is not straightforward. For hybrid kV/MV image-guided radiotherapy imaging, larger diameter markers are suggested. If using kV imaging alone, smaller sized markers may be used in smaller sized patients in order to reduce artifacts. Only larger diameter gold markers are visible across all imaging modalities.
评估用于放射治疗计划和治疗引导的各种商业标记物在磁共振成像(MRI)、计算机断层扫描(CT)和超声成像中的可视性、伪影和畸变情况。
我们比较了来自3个供应商的2种实心金标记物、4种金线圈和1种聚合物标记物。使用的成像模态包括3-T和1.5-T的GE MRI、西门子Sequoia 512超声、飞利浦大孔径CT、瓦里安Trilogy直线加速器(锥形束CT [CBCT]、机载成像器千伏 [OBI-kV]、电子射野影像装置兆伏 [EPID-MV])以及美敦力O-ARM CBCT。标记物在一个30×30×10 cm³的定制团注体模中成像。在一个实验中,在标记物周围使用外科用润滑剂以减少气隙。图像以医学数字成像和通信(DICOM)格式保存,并使用内部软件进行分析。标记物的剖面用于对标记物信号进行客观比较。对每个标记物产生的可视性和伪影/畸变进行定性和定量评估。
所有标记物在CT、CBCT、OBI-kV和超声中均可见。直径小于0.75 mm的金标记物在EPID-MV图像中不可见。标记物越大,CT和CBCT图像伪影越多,但其伪影程度取决于扫描参数和扫描仪本身。在所研究的所有成像模态中,直径0.75 mm及以上的金线圈的可视性相当。聚合物标记物在CT和CBCT中产生的伪影最小,但在EPID-MV中的可视性较差。0.5 mm的金线圈由于尺寸小,在MRI和EPID-MV中的可视性较差。根据扫描序列,金标记物在3-T T1梯度回波序列中比在1.5-T T1快速自旋回波序列中更易见。在本研究中,所有标记物在超声上均清晰可见。
所有金标记物在CT、CBCT、千伏成像和超声中均可见;然而,只有大直径标记物在兆伏成像中可见。当使用MR和EPID-MV成像器时,基准标记物的选择并不简单。对于混合千伏/兆伏图像引导放射治疗成像,建议使用较大直径的标记物。如果仅使用千伏成像,对于体型较小的患者可使用较小尺寸的标记物以减少伪影。只有较大直径的金标记物在所有成像模态中均可见。