四种不同晶状体测量仪的比较。

A comparison of four different lens mappers.

作者信息

Larrue Denis, Legeard Morgane

机构信息

*PhD †MS Laboratoire Alutec, Morez, France (both authors).

出版信息

Optom Vis Sci. 2014 Nov;91(11):e260-6. doi: 10.1097/OPX.0000000000000396.

Abstract

PURPOSE

Recently, a number of lens mappers have become available for measuring the detailed optical properties of progressive addition lenses (PALs). The goal of this study was to compare the results obtained from several different lens mappers for a range of different lenses.

METHODS

The optical power maps of six lenses-two single-vision lenses, a parallel-sided slide, a flat prism, and two progressive lenses-were measured using four different lens mappers: the Dual Lens Mapper, the Nimo TR4005, the Rotlex Class Plus, and the Visionix VM2500. The repeatability of the instruments was also evaluated.

RESULTS

All lens mappers gave very repeatable measurements; however, measurements among the lens mappers varied considerably. Differences appeared to be above the tolerance at the optical center for measurements of single-vision lenses, and these differences increase in the periphery up to 1.00 diopter. Similar differences were observed for the PALs, even increased by prism and base curve effect, with figures greater than 1 diopter in the periphery. The measurements made on the prism and lenses with different base curves suggest that base curve, thickness, and prismatic effect can all contribute to the differences among instruments.

CONCLUSIONS

Measurements of a given lens taken with different lens mappers can vary substantially. Particular caution should be exercised when interpreting power maps for PALs taken with different instruments.

摘要

目的

最近,有多种镜片测绘仪可用于测量渐进多焦点镜片(PAL)的详细光学特性。本研究的目的是比较几种不同镜片测绘仪对一系列不同镜片的测量结果。

方法

使用四种不同的镜片测绘仪:双镜片测绘仪、Nimo TR4005、Rotlex Class Plus和Visionix VM2500,测量了六个镜片的屈光力图,其中包括两个单焦点镜片、一个平行边滑块、一个平面棱镜和两个渐进多焦点镜片。还评估了这些仪器的重复性。

结果

所有镜片测绘仪的测量结果都具有很高的重复性;然而,不同镜片测绘仪之间的测量结果差异很大。对于单焦点镜片的测量,在光学中心处的差异似乎超出了公差范围,并且这些差异在外围增加到1.00屈光度。对于渐进多焦点镜片也观察到了类似的差异,甚至由于棱镜和基弧效应而有所增加,在外围的差异大于1屈光度。对棱镜和具有不同基弧的镜片进行的测量表明,基弧、厚度和棱镜效应都可能导致仪器之间的差异。

结论

使用不同的镜片测绘仪对给定镜片进行测量时,结果可能会有很大差异。在解释使用不同仪器拍摄的渐进多焦点镜片的屈光力图时应格外谨慎。