Suppr超能文献

在NIMO TR1504(Lambda-X)上测量的屈光度分布的可靠性以及单焦点、双焦点和多焦点隐形眼镜的镜片偏心效应。

Reliability of power profiles measured on NIMO TR1504 (Lambda-X) and effects of lens decentration for single vision, bifocal and multifocal contact lenses.

作者信息

Kim Eon, Bakaraju Ravi C, Ehrmann Klaus

机构信息

Brien Holden Vision Institute, Sydney, Australia.

Brien Holden Vision Institute, Sydney, Australia.

出版信息

J Optom. 2016 Apr-Jun;9(2):126-36. doi: 10.1016/j.optom.2015.10.005. Epub 2015 Nov 14.

Abstract

PURPOSE

To evaluate the repeatability of power profiles measured on NIMO TR1504 (Lambda-X, Belgium) and investigate the effects of lens decentration on the power profiles for single vision (SV), bifocal (BF) and multifocal (MF) contact lenses.

METHODS

Accuracy of the sphere power was evaluated using single vision BK-7 calibration glass lenses of six minus and six plus powers. Three SV and four BF/MF contact lenses - three lenses each, were measured five times to calculate the coefficients of repeatability (COR) of the instrument. The COR was computed for each chord position, lens design, prescription power and operator. One lens from each type was measured with a deliberate decentration up to ±0.5mm in 0.1mm steps.

RESULTS

For all lenses, the COR varied across different regions of the half-chord position. In general, SV lenses showed lower COR compared to the BF/MF group lenses. There were no noticeable trends of COR between prescription powers for SV and BF/MF lenses. The shape of the power profiles was not affected when lenses were deliberately decentered for all SV and PureVision MF lenses. However, for Acuvue BF lenses, the peak to trough amplitude of the power profiles flattened up to 1.00D.

CONCLUSION

The COR across the half-chord of the optic zone diameter was mostly within clinical relevance except for the central 0.5mm half-chord position. COR were dependent on the lens type, whereby BF/MF group produced higher COR than SV lenses. The effects of deliberate decentration on the shape of power profiles were pronounced for lenses where the profiles had sharp transitions of power.

摘要

目的

评估在NIMO TR1504(比利时Lambda-X公司)上测量的屈光力分布的可重复性,并研究镜片偏心对单焦点(SV)、双焦点(BF)和多焦点(MF)隐形眼镜屈光力分布的影响。

方法

使用六种负度数和六种正度数的单焦点BK-7校准玻璃镜片评估球镜屈光力的准确性。对三片SV和四片BF/MF隐形眼镜(每种三片)各测量五次,以计算仪器的可重复性系数(COR)。针对每个弦位置、镜片设计、处方屈光力和操作人员计算COR。对每种类型的一片镜片进行故意偏心测量,偏心量最大为±0.5mm,步长为0.1mm。

结果

对于所有镜片,COR在半弦位置的不同区域有所变化。总体而言,与BF/MF组镜片相比,SV镜片的COR较低。SV和BF/MF镜片的处方屈光力之间没有明显的COR趋势。对于所有SV和PureVision MF镜片,当镜片故意偏心时,屈光力分布的形状不受影响。然而,对于Acuvue BF镜片,屈光力分布的峰谷幅度变平达1.00D。

结论

除了光学区直径的中心0.5mm半弦位置外,光学区直径半弦上的COR大多在临床相关范围内。COR取决于镜片类型,BF/MF组产生的COR高于SV镜片。对于屈光力分布有急剧变化的镜片,故意偏心对屈光力分布形状的影响较为明显。