Talanta. 2015 Jan;131:379-85. doi: 10.1016/j.talanta.2014.08.024. Epub 2014 Aug 19.
An electrochemiluminescence (ECL) immunosensor using Pt electrode modified with carbon quantum dot (CQDs) coated Au/SiO2 core-shell nanoparticles was proposed for sensitive detection of 8-hydroxy-2'-deoxyguanosine (8-OHdG) in this work. Rabbit anti-8-OHdG antibody was covalently bound to CQDs on the surface of Au/SiO2 core-shell nanoparticles. Through signal amplification of Au/SiO2 core-shell nanoparticles, 8-fold enhancement of the ECL signals was achieved. Under optimal conditions, a good linear range from 0.2 to 200 ng mL(-1) with a low detection limit of 0.085 ng mL(-1) (S/N=3) for 8-OHdG detection was obtained. Interfering substances tests showed that the corresponding ECL intensity (ΔECL) of 8-OHdG is 8-18 times higher than that of guanine, uric acid (UA) and ascorbic acid, demonstrating its good selectivity for 8-OHdG detection. The ECL immunosensor exhibits long-term stability with a relative standard deviation (RSD) of 8.5% even after 16 cycles of continuous potential scans. The result of analytical detection of 8-OHdG in real samples was satisfactory. The proposed ECL immunosensor shows good performance with high sensitivity, specificity, repeatability, stability and provided a powerful tool for 8-OHdG monitoring in clinical samples.
本文提出了一种基于 Pt 电极修饰的碳量子点(CQDs)包覆的 Au/SiO2 核壳纳米粒子的电化学发光(ECL)免疫传感器,用于灵敏检测 8-羟基-2'-脱氧鸟苷(8-OHdG)。兔抗 8-OHdG 抗体通过共价键结合到 Au/SiO2 核壳纳米粒子表面的 CQDs 上。通过 Au/SiO2 核壳纳米粒子的信号放大,实现了 ECL 信号的 8 倍增强。在最佳条件下,获得了 0.2 至 200ng mL(-1)的良好线性范围,检测限低至 0.085ng mL(-1)(S/N=3)。干扰物质测试表明,8-OHdG 的相应 ECL 强度(ΔECL)比鸟嘌呤、尿酸(UA)和抗坏血酸高 8-18 倍,表明其对 8-OHdG 检测具有良好的选择性。ECL 免疫传感器具有长期稳定性,即使在连续电位扫描 16 个循环后,相对标准偏差(RSD)仍为 8.5%。实际样品中 8-OHdG 的分析检测结果令人满意。该 ECL 免疫传感器具有高灵敏度、特异性、重复性、稳定性,为临床样品中 8-OHdG 的监测提供了有力工具。