Fajardo Emmanuel, Metcalf Carol, Mbofana Elton, van Vyve Charlotte, Munyaradzi Dhodho, Simons Sandra, Kuhudzayi Misheck, Bygrave Helen
BMC Infect Dis. 2014 Oct 22;14:553. doi: 10.1186/s12879-014-0553-9.
CD4+ T-cell testing of blood specimens collected in standard EDTA Vacutainer tubes and transported at ambient temperature, must be completed within 48 hours with the BD FACSCount™ flow cytometer, restricting specimen collection in remote clinics with no on-site testing and limited specimen transport services. We conducted a study in Buhera District, Zimbabwe, to assess the stability and accuracy of CD4+ T-cell results of samples collected in Stabilization Tubes (ST) and stored at ambient temperature for varying time periods.
Paired EDTA and ST samples were collected from 51 HIV-positive patients aged 18 years and older. CD4+ T-cell testing was done on arrival in the laboratory (Day 0). ST samples were retested on Days 3, 5, and 7. Nineteen ST samples were stored for an additional week and retested on Day 14.
There was a strong correlation between absolute CD4+ T-cell counts measured in the EDTA Day 0 reference sample and Day 7 ST sample (Spearman's rho: 0.9778; mean difference: -4.9 cells/μL and limits of agreement (LOA): 98.5 and 88.7 cells/μL); and the reference sample and Day 14 ST sample (Spearman's rho: 0.9632; mean difference 5.1 cells/μL and LOA: -99.6 and 109.8 cells/μL. Using a 350 cells/μL threshold, the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were all 100% on Day 7, and 83.3%, 100%, 100% and 92.9% on Day 14. Using a 500 cells/μL threshold, the sensitivity, specificity, PPV and NVP were 100%, 88.5%, 88.5% and 100% on Day 7 and 88.9%, 80.0%, 80.0% and 88.9% on Day 14.
CD4 ST can be used and stored up to 7 days as a reliable alternative to standard EDTA tubes in settings where CD4+ T-cell testing within 48 hours is not feasible. Despite the small sample size, results suggest that ST may be stored up to 14 days at room temperature for CD4 testing, without compromising accuracy. However, further studies with larger sample sizes are needed to confirm this preliminary finding.
使用BD FACSCount™流式细胞仪对采集于标准乙二胺四乙酸(EDTA)真空采血管并在常温下运输的血液标本进行CD4+ T细胞检测时,必须在48小时内完成,这限制了在没有现场检测且标本运输服务有限的偏远诊所进行标本采集。我们在津巴布韦的布赫拉区开展了一项研究,以评估采集于稳定化管(ST)并在常温下保存不同时间段的样本中CD4+ T细胞检测结果的稳定性和准确性。
从51名18岁及以上的HIV阳性患者中采集配对的EDTA和ST样本。到达实验室当天(第0天)进行CD4+ T细胞检测。ST样本在第3天、第5天和第7天重新检测。19个ST样本额外保存一周,并在第14天重新检测。
在第0天的EDTA参考样本和第7天的ST样本中测得的绝对CD4+ T细胞计数之间存在强相关性(斯皮尔曼等级相关系数:0.9778;平均差异:-4.9个细胞/微升,一致性界限(LOA):98.5和88.7个细胞/微升);以及参考样本和第14天的ST样本之间(斯皮尔曼等级相关系数:0.9632;平均差异5.1个细胞/微升,LOA:-99.6和109.8个细胞/微升)。使用350个细胞/微升的阈值,第7天时敏感性、特异性、阳性预测值(PPV)和阴性预测值(NPV)均为100%,第14天时分别为83.3%、100%、100%和92.9%。使用500个细胞/微升的阈值,第7天时敏感性、特异性、PPV和NVP分别为100%、88.5%、88.5%和100%,第14天时分别为88.9%、80.0%、80.0%和88.9%。
在无法在48小时内进行CD4+ T细胞检测的情况下,CD4稳定化管可作为标准EDTA管的可靠替代品使用并保存长达7天。尽管样本量较小,但结果表明,稳定化管在室温下保存长达14天用于CD4检测时,不会影响准确性。然而,需要进一步开展更大样本量的研究来证实这一初步发现。